Thermocouple Ranges & Accuracies
Table based on Accuracy:
≤ ± (0.008 % of Reading + 0.006 mV)
Note: Doesn't include cold junction error of ±0.05°C
T/C
Degrees C
°C
Range
R
-18.30 to 250.00
±1.26°
250.00 to 750.00
±0.64°
750.00 to 1600.00
±0.54°
1600.00 to 1767.80
±0.63°
S
-18.30 to 150.00
±1.22°
150.00 to 500.00
±0.72°
500.00 to 1650.00
±0.63°
1650.00 to 1767.80
±0.73°
B
315.60 to 550.00
±1.88°
550.00 to 900.00
±1.03°
900.00 to 1150.00
±0.72°
1150.00 to 1820.00
±0.63°
N
-230.00 to -100.00
±1.10°
-100.00 to 0.00
±0.30°
0.00 to 1100.00
±0.24°
1100.00 to 1300.00
±0.27°
Page 43
GlobalTestSupply
www.
Degrees F
°F
Range
-1.00 to 482.00
±2.27°
482.00 to 1382.00
±1.14°
1382.00 to 2192.00
±0.97°
2192.00 to 3214.00
±1.13°
-1.00 to 302.00
±2.20°
302.00 to 932.00
±1.30°
932.00 to 3002.00
±1.14°
3002.00 to 3214.00
±1.31°
600.00 to 1022.00
±3.39°
1022.00 to 1652.00
±1.86°
1652.00 to 2102.00
±1.30°
2102.00 to 3308.00
±1.14°
-382.00 to -148.00
±1.98°
-148.00 to 32.00
±0.53°
32.00 to 2012.00
±0.44°
2012.00 to 2372.00
±0.49°