Warnings - Verity Medical NeuroTrac Simplex Operation Manual

Single channel emg
Table of Contents

Advertisement

Warnings

*
This unit must be used with the guidance of a Physiotherapist or Doctor.
*
Type BF equipment, Continuous Operation.
*
Do not insert lead wires into a main power supply.
*
Do not immerse in water or any other substance.
*
The unit is not protect from the ingress of water droplets from a shower
of rain if used outside the carrying case.
*
Do not use the NeuroTrac
anaesthetic gas mixture and air or with Oxygen or Nitrous Oxide.
*
If using rechargeable 9 Volt PP3 Nickel Metal Hydride batteries, be
sure to use a CE approved battery charger. Never connect the
NeuroTrac
mains powered equipment.
We advise not to use Ni-Cad rechargeable batteries.
*
To avoid the effects of electromagnetic interference never use the
NeuroTrac
other powerful radio interference producing equipment that produces
electrical sparks etc. In the EMG Mode, the NeuroTrac
be susceptible to strong interfering radio type emissions that may lead
to temporally increased EMG microvolt readings. The reading will
immediately return to the correct value when the interference ceases.
(Remember that a relaxed muscle should read below 3.5 µVolts).
*
Patient Electrodes including all skin surface electrodes Vaginal
Electrodes and rectal probe are for Single Patient Use only.
*
Keep out of reach of children.
*
No modification of this equipment is allowed!
®
Simplex directly to a battery charger or to any other
®
Simplex, within 4 metres of a cellular telephone or near any
®
NeuroTrac
Simplex - Operation Manual
®
Simplex in the presence of a flammable
4
Simplex may
®