Emg Session Settings - Verity Medical NeuroTrac Simplex Operation Manual

Single channel emg
Table of Contents

Advertisement

EMG Session settings

To adjust the session parameters, press and hold SET button for 3 seconds.
The first parameter will be VOL (sound volume). To adjust the value, press
+/- buttons. To go to the next parameter in the table below, press SET button.
To save new settings, simply press ESC or OFF. The new settings will be
immediately stored.
EMG
parameters
Sound Volume
Work Time
Rest Time
Number of
Trials
Audio
Biofeedback
type
PC Data
output
EMG Notch
Filter
Text on
LCD
Value range: 0 - sound off, 1,2,3,4,5 - quiet, 5,6,7,8,9,10
- loud.
Typical value: 4.
VOL
This is the sound volume of beeps and prompts: Work,
Rest, Lock, PC-CONNECT, Erase
Value range: [sec]: 2,3,4, ..., 99
Typical value: 5.
WRK
This is the time during which the patient is prompted
to keep the measured muscles contracted above the
threshold level.
Value range: [sec]: 2,3,4, ..., 99.
Typical value: 5.
RST
This is the time during which the patient is prompted to
keep the measured muscles relaxed below the threshold
level, ideally below 1uV.
Value range: 2,3,.., 99.
Typical value: 5.
TRS
This parameter defines how many repetitions of Work/
Rest cycles your treatment requires.
Select from the following:
FAB - the EMG beeps will be audiable only above the
FAB
threshold.
FBL
FBL - the EMG beeps will be audiable only below the
FCO
threshold.
FOF
FCO - the EMG beeps will be audiable above and
below the threshold
FOF - no EMG beeps.
Select from the following:
DON
DON: - switch ON to enable PC Software connection.
DOF
DOF: - switch OFF to save batteries.
Typical setting: DON.
Select from the following:
WDE
(typical setting)
be more sensitive but it will also pick up heart beats.
WDE
NRW - narrow filter - your EMG reading will be filtered
NRW
to reduce the heart beat spikes on the EMG reading. Use
narrow filter if your electrodes are in close proximity to
heart (on chest, back, arms, vaginally or rectally).
NeuroTrac
Simplex - Operation Manual
®
Available settings
- wide filter - your EMG reading will
15
Typical setting: FAB.