Table 204:
The NeuroTrac® product is intended for use in the electromagnetic environment
specified below. The customer or the user of the NeuroTrac® product should as-
sure that it is used in such an environment.
Immunity
IEC 60601
test
test level
3 Vrms
150 kHz to
Conducted RF
80 MHz
IEC 61000-4-6
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2,5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic en-
vironment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which NeuroTrac® product is used exceeds
the applicable RF compliance level above, the NeuroTrac® product should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the NeuroTrac® product.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
NeuroTrac
Guidance and manufacturer's declaration
– electromagnetic immunity
Compliance
level
3 Vrms
150 kHz to 80
MHz
3 V/m
80 MHz to 2,5
GHz
27
Simplex - Operation Manual
®
Electromagnetic environment
guidance
Portable and mobile RF communications
equipment should be used no closer
NeuroTrac®
to any part of the
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
d = 1.2 √P 150 kHz to 80 MHz,
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5GHz
where P is the maximum output power
rating of the transmitter in watts (W) ac-
cording to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmit-
ters, as determined by an electromag-
a
netic site survey
, should be less than
the compliance level in each frequency
range b .
Interference may occur in the vicinity of
equipment marked
with the following
symbol:
product,