Specifications - Verity Medical NeuroTrac Simplex Operation Manual

Single channel emg
Table of Contents

Advertisement

Specifications

EMG Electrical Specification
1.
Single Channel
2.
EMG Range: 0.2 to 2000 µV RMS (continuous)
3.
Sensitivity: 0.l µV RMS
4.
Accuracy: 4% of µV reading +/-0.3 µV at 200 Hz
5.
Selectable Bandpass filter - 3db Bandwidth,
a. Wide:
b. Narrow: 100 Hz +/- 5% to 370 Hz +/- 10%
6.
Notch filter: 50 Hz (Canada 60Hz) - 33 dbs (0. 1% accuracy)
7.
Common Mode Rejection Ratio: 130 dbs Minimum @ 50 Hz
8.
Work / Rest periods: 2-99 seconds
9.
Number of Trials: 1-99
10.
Battery: PP3 Alkaline, 9V.
Expected average battery life [of standard 800 mAh, alkaline]:
40 hours.
11.
Low battery indication at 7.4 volts +/- 0.2 volts and automatic shut off
10 minutes after last key pressed, unless infra-red is turned on.
Expected service life:
5 years. Careful use and maintenance extends the life of the unit over the
service life limit.
Environmental Conditions for use:
+5 to +40 degrees Centigrade. 15-93% Humidity.
Environmental conditions for storage & transport:
-25 to +70 degrees Centigrade. 15-93% Humidity.
Physical Dimensions:
Length 128.5mm, Width 64mm, Depth 28.3 mm excluding belt clip
Weight: 150g.
18 Hz +/- 4 Hz to 370 Hz +/- 10%
- Reading below 235 microvolts
10 Hz +/-3 Hz to 370 Hz +/- 10%
- Reading above 235 microvolts