Memory Location - GE ERESCO MF4 Series Operating Manual

Industry x-ray unit for digital control unit
Table of Contents

Advertisement

14 Memory location
The memory location defines the location (data source) where all relevant system data
including operating history, warm-up history, program list and configuration data) is
centrally adminstrated.
The memory location can be changed by pressing the F2   key.
If the active data source is the ERESCO MF4 control, all history events will be logged in the
control module and radiation unit. However, upon access of this information (download,
display) only the data stored in the control will be used.
The program list including the program names will only be managed in the control module.
If the active data source is the X-ray unit, the control module does not log or store any
data. Upon data access (download, display) the control module accesses the data source
on-board of the X-ray unit and temporarily downloads it.
This mode is downwards compatible to MF3 operations.
This mode does not support naming conventions for exposure programs.
By using the key F3
overwritten with the data that is stored in the X-ray unit (e.g. for field service).
Note:
Copyright by GE Sensing & Inspection Technologies GmbH
08/24/2016
6. Edition, ID No.: 37020GB-F
Operating Manual ERESCO MF4 Type Series
, the data which is stored in the ERESCO MF4 control will be
If the Administrator Software program is present, the data can be administrated if
necessary with this program.