04 Prewarning

The current prewarning time is displayed.
Press the F5
Press the F2
prewarning time inversely:
This setting will be changed by entering another value via the numeric keypad or by pressing
the  or  keys.
Press the OK key to confirm the new setting.
Enter 0 to switch off the prewarning time.
After entering 0 second, the password 1483 must be entered to confirm the setting.
Password input is not required for prewarning time changes above 2 seconds. A prewarning
time setting of 1 second is not possible.
CAUTION:
NOTE:
The prewarning time is factory set to 2 seconds.
Copyright by GE Sensing & Inspection Technologies GmbH
08/24/2016
6. Edition, ID No.: 37020GB-F
Operating Manual ERESCO MF4 Type Series
key to exit the menu item without changes.
key to change the function key arrangement and to display the current
The prewarning time is designed to protect personnel and should not be
set to 0!
With a connected warning flash lamp, an prewarning time of at least 2
seconds m u s t be set, otherwise the high voltage cannot be switched on.