NOTE: Ramp Time must be set to a minimum of 5 minutes prior to
attempting to program Ramp Start Pressure or the change to Ramp Start
Pressure will not take effect and "FAIL 2" will be shown in the General Status
Display field .
Setting Ramp Start Pressure
Press and hold the Down Arrow Button for 3 seconds until the ramp start pressure value flashes .
While flashing, repeatedly press either the Up or Down Arrow Buttons to increase or decrease the
ramp start pressure by 0 . 5 cm H
O increment . (Figure 12) Ramp start pressure must be lower than
2
prescribed pressure .
When the desired value for ramp start pressure is reached, hold down the Start/Stop Button for
approximately 3 seconds to save your setting . The Z1 will return to standby mode . To return to
standby mode without saving the new setting, press the Start/Stop Button momentarily instead
of holding it down .

User Software

Human Design Medical provides User Software for data viewing which may be downloaded from
the Human Design Medical website to any Windows computer . A separate user guide for this
software is available for download at the Human Design Medical website, www . h dmusa . c om . This
software can be used to upgrade the Z1 firmware as new releases are made available .

Charging the Battery (PowerShell Only)

WARNING: Only charge the Battery with the
supplied Human Design Medical PowerShell
and the supplied Human Design Medical Power
Supply. Do not use any other battery charger, AC
adapter, or other device to charge the Battery .
With both the Z1 and the Battery fully inserted into the
PowerShell, plug the Power Supply into the PowerShell as shown
and into a power outlet . (Figure 13)
CAUTION:
Make sure your Battery is fully
charged before commencing CPAP therapy using
the PowerShell . The battery life of a fully charged
Extended Life Battery in the PowerShell is typically
8 . 9 hours or longer, after which treatment will
terminate .
Ensure the Z1 is in standby mode (display reads 0 . 0 ) . The Battery
will not charge unless the Z1 is in standby mode . The Battery
status indicator on the Z1 will flash when the Battery is charging .
The number of solid bars indicates the current level of charge .
(Figure 14)
The battery status indicator will stop flashing when the Battery
is fully charged . Typically it will take 12 hours to fully charge the battery . After a few minutes, the
backlit display will turn off . This is to conserve power . It does not indicate that the battery is fully
charged .
16
NOTE: The battery will not charge if you are using your Z1 CPAP at the
same time .

Care

The Z1 System needs periodic care . It is the user's responsibility to be aware of the following items
for care:

• Inspections

• Cleaning

• Filter Replacement
• CPAP Tube Replacement
• Reminders and Fault Conditions
Inspections
Before each use carefully inspect the Z1, CPAP Tube, Filter, and mask for signs of wear, damage or
discoloration . Contact your retail medical supplier or at www . h dmusa . c om for replacement parts .
Cleaning
Regular cleaning should be performed in accordance with the instructions below .
immerse the Z1, Power Adapter, Power Cord or Battery in water . Always
unplug these devices before cleaning (only as instructed in this User Guide)
and make sure they are dry before plugging back in .
CAUTION:
solutions, moisturizing or antibacterial soaps or scented oils to clean the
Z1 . Never attempt to clean the Power Adapter, PowerShell, Battery or
accessories with any of these solutions . These solutions may create a shock
hazard, cause damage and reduce the life of these products .
Figure 13
Daily
1 . Remove the CPAP Tube and the mask from the Z1 .
2 . Wash the CPAP Tube in warm water using mild detergent .
3 . Rinse thoroughly, hang, and allow to dry .
4 . Before next use, reconnect the CPAP Tube and the Tube Adapter to the Z1 and mask .
NOTE: Soaking the CPAP Tube in water for longer than 10 minutes may
shorten the life of the tube .
Figure 14
WARNING:
This device can expose you to electrocution . Do not
Do not use bleach, chlorine, alcohol, or aromatic-based
17