Items and components not manufactured and resold by DynAmp, LLC are warranted by their manufacturer. Warranty repair shall be, at DynAmp’s option, in the form of repair or replacement of the defective items or components. Concerning warranty repairs, DynAmp, LLC will be responsible for DynAmp provided time, material and transportation costs (shipping or travel).
DynAmp, LLC does not assume liability for the customer’s failure to comply with the rules and requirements provided in this manual. This equipment is designed to be connected to hazardous electric voltages.
If a binding screw is used for the mains earth connection, the binding screw must be size M4 (No. 6) or larger. DynAmp, LLC does not assume liability for the customer’s failure to comply with the rules and requirements provided in this manual.
LKAT Installation, Operation and Service Manual +24V DynAmp, LLC Customer Support & Service Assistance For further assistance, contact DynAmp Customer Support at: Americas: Telephone: +1 614.871.6900 Fax: +1 614.871.6910 8:00 AM to 5:00 PM USA Eastern Time From first Sunday in November to second Sunday in March – 13:00 GMT to 22:00 GMT From second Sunday in March to first Sunday in November –...
Page DynAmp, LLC WARRANTY ________________________________________________ iii Hazard Warning! _________________________________________________________ v SAFETY _______________________________________________________________ vii DynAmp, LLC Customer Support & Service Assistance ________________________ ix MANUAL REVISIONS _____________________________________________________ xi TABLE OF CONTENTS ___________________________________________________ xiii 1. HANDLING AND STORAGE______________________________________________________ 1 2. PRODUCT DESCRIPTION _______________________________________________________ 3 2.1 SYSTEM OVERVIEW ......................
The LKAT combines accurate high current measurement with reverse and overcurrent protection in a single rugged and cost effective system. The next generation of DynAmp’s well proven OLOP technology, LKAT offers a new level of accuracy and stability to accurately and reliably measure uni- and bi-directional DC as well as AC bus currents.
Checking field calibration of LKAT Systems requires use of a reference current transducer with calibration accuracy of 0.1% full-scale or better. Contact DynAmp, LLC for calibration of the LKAT System. Refer to “Maintenance and Spare Parts” section for more information.
(refer to Figure 3.1 for additional info.) Temperature Sensitivity ±50ppm/°C or better Mains Voltage Sensitivity ±0.001%/V * At DynAmp reference conditions : Ambient 25°C ± 2°C (77°F ± 4°F) / Mains 18 – 36Vdc MAIN STATUS INDICATORS Accuracy Diagnostics Indicates LKAT System proper operation or...
The complete system should be inspected for shipping damage at the earliest opportunity. Visible damage must be reported to the carrier immediately. Concealed damage (not evident until the system is operated) must be reported to DynAmp, LLC immediately. INSTALLATION CONSIDERATIONS Read the following before mounting the head: 1.
Laboratory Accreditation Bureau "Guidance for Documenting and Implementing ISO/IEC 17025:2005 and Laboratory Guidance." As a guideline, DynAmp recommends a 24-month interval of calibration for all permanently installed products and a 12-month interval of calibration for all products used in portable applications.
Field calibration is possible using a 0.1% accurate current measurement system, such as the DynAmp LKP, LKCO or Opti-Cal as a reference. Due to significantly larger size and mass, it may be difficult to install a comparably rated LKP Measuring Head in on the same bus as the LKAT .
The trip setpoint is proportional to the current where the trip will occur: “Trip Current” – from Metering Unit nameplate. The easiest method to check or change trip setpoints is to use the DynAmp LKAT MUT system which is specifically designed to simplify this process as well as verify proper operation afterwards.
PERIODIC MAINTENANCE As is true with any electronic system, proper maintenance will tend to prolong the service life. DynAmp, LLC recommends the following program be performed at the recommended interval to prevent or detect damage to the LKAT System and to ensure reliable performance.
System Measuring Heads require special repair procedures and materials. However, the heads are very reliable and should require little or no repair over its service life. Please refer to the nearest authorized DynAmp, LLC service center for information on repairs. Table 6.1...