Measuring Parameter; Description - BAQ alphaDUR mini Manual

Hide thumbs Also See for alphaDUR mini:
Table of Contents

Advertisement

MEASURING PARAMETER

6 Measuring parameter

6.1 Description

After power-up, the alphaDUR mini restores the last used measuring
parameter.
The measuring parameters in detail:
M a t e r i a l :
H a r d n e s s s c a l e : The current hardness scale (Vickers (HV), Rockwell (HRC
R a t i n g :
S t a t i s t i c s :
I n s t a n t p r i n t o u t : If a mobile printer is connected, a consecutive printout can
20
The currently selected material calibration. For details
about material calibration see chapter 8.
or HRB), Brinell (HB) or tensile strength (N/mm
which the readings will be converted. The result of an UCI
measurement is Vickers hardness. If another hardness
scale is selected, the values will be converted according
to DIN standard 50150. In the measurement window the
hardness scale can be switched with the aid of the key
SCALE, unless the measured value memory is activated.
These are the upper and lower limits for rating. A signal of
2 short beeps sounds if a measured value lies outside
these limits. A single beep sounds if the value is rated
GOOD.
If both limits are set to 0, no rating will be done.
It is self-evident, that the the upper limit must be higher
than the lower limit.
If rating is activated, in will be indicated in the measuring
window when the upper or lower limit is exceeded.
The number of measured values which should be
statistically evaluated without the use of the measured
value memory is defined here.
be made (see 5.6). This parameter can be set On or Off.
2
)), to

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents