Manual
Version 1.2

Advertisement

Table of Contents
loading

Summary of Contents for BAQ alphaDUR mini

  • Page 1 Manual Version 1.2...
  • Page 3: Table Of Contents

    Contents 1 Introduction....................5 2 Manual.......................6 2.1 Lettering....................6 2.2 Text boxes...................6 3 The test probes..................7 3.1 Probe selection...................7 3.2 Probe handling..................7 4 General working instructions..............9 4.1 Key functions..................9 4.2 Status bar....................10 4.3 The menus..................10 4.4 Text entry.....................10 4.5 The number field.................12 5 Measurement.....................13 5.1 Specimen requirements..............13 5.2 Measuring parameter................14...
  • Page 4 9.2 Continuation of a series..............27 9.3 Deleting a series.................27 9.4 Displaying a series................27 9.5 Optional: Copying of series to an USB flash drive......27 10 System settings..................28 10.1 Language..................28 10.2 Time....................28 10.3 Date....................28 10.4 Configuration..................28 10.4.1 Date format.................28 10.4.2 Key SCALE.................28 10.4.3 Key MAT..................29 10.4.4 Query whether series should be saved........29 10.4.5 Query whether values should be printed........29...
  • Page 5: Introduction

    50150. Up to 100 000 readings including hardness, date, time and measuring parameter can be stored to the internal memory of the alphaDUR mini. The memory can be divided into sections for use by different users or projects. Saved series of measurements, including statistics, can be readily displayed or printed.
  • Page 6: Manual

    MANUAL 2 Manual 2.1 Lettering Operating elements Operating elements Keys Standard text Descriptions and explanations in text form Important notes Important notes and remarks 2.2 Text boxes Explanations of functions and operating Operating elements elements are framed in blue. Green fields describe user inputs and Entry fields parameters Information and notes are provided with yellow...
  • Page 7: The Test Probes

    THE TEST PROBES 3 The test probes 3.1 Probe selection alphaDUR UCI test probes are available with test loads of 10, 20, 30, 49, and 98 N. This corresponds to HV1, HV2, HV3, HV5 and HV10 (1, 2, 3, 5 and 10 kg). So the optimal load for a particular test task can be chosen.
  • Page 8 THE TEST PROBES ial and must be in measuring mode.) The Vickers diamond may touch the surface slightly, but not for too long (otherwise an error occurs). Then the probe is pressed to the specimen until the protective sleeve hits the sur- face.
  • Page 9: General Working Instructions

    GENERAL WORKING INSTRUCTIONS 4 General working instructions 4.1 Key functions Switching on/off the device. Changes the hardness scale. Changes the material. Delete last measurement point Display statistics Into the main menu Cursor keys With these keys, menu items are selected and the re- quired values in input fields fields are set.
  • Page 10: Status Bar

    GENERAL WORKING INSTRUCTIONS 4.2 Status bar The rechargeable battery status and the time are displayed in the status bar at the top of the window. 4.3 The menus A menu consists of a list of available menu items and a bar which identifies the active menu item.
  • Page 11 GENERAL WORKING INSTRUCTIONS Switching between upper case and lower case Accepting text and closing text entry window Cancel Closing text entry window without accepting the text The field on which the focus is placed is identified by a yellow background. With the keys ▲...
  • Page 12: The Number Field

    GENERAL WORKING INSTRUCTIONS 4.5 The number field A number field is used for the input of numbers. It consists of mostly several digits which can be changed separately, and a marker (cursor) which can be displaced with the cursor keys ◄ and ► within the number field. The Item on which the cursor is at this time can be modified through pressing the cursor keys ▲...
  • Page 13: Measurement

    5 Measurement The measurement can be started with the menu item Measurement. If a probe is connected to the alphaDUR mini, the measurement window will by shown directly after start-up. The test load of the connected probe will be detected automatically.
  • Page 14: Measuring Parameter

    MEASUREMENT To avoid this effect, the specimen has to be thick enough to allow the os- cillations to attenuate before the diamond is reached. With thin speci- men, the mass of the specimen has an effect. If the mass of the speci- men is high enough, a thickness of 8 mm for flat specimen or 10 mm for round stock is sufficient.
  • Page 15: Measuring Procedure

    MEASUREMENT meets the requirement already has been saved, you can use the menu item Measuring parameter/Load (see 6.2.3) to reload them. After power-up, the alphaDUR mini restores the measuring parameter which were last used. 5.3 Measuring procedure To take a measurement, put the probe vertically onto the specimen and press it down steadily as far as it will go.
  • Page 16: Key Assignment Of The Measuring Window

    MEASUREMENT Charge status display:: Displays the residual charge of the batteries. Time: Current time Measured value: Current measured value Hardness scale: Current hardness scale Material: Current material Number of Number of measurements already carried out measurements: Average value: Average value of the measurements carried out Standard deviation: Standard deviation History:...
  • Page 17: Statistics

    MEASUREMENT 5.5 Statistics If the measured value memory is switched on, the values of a series can be evaluated statistically at all times. Even if the measured value memory is not switched on, the hardness numbers are always temporarily stored until the measurement window is closed.
  • Page 18 STAT or because a value has been deleted) the measurement series will be continued. The alphaDUR mini can be configured to display a query whether the data should be saved as a new series (see 10.4.4) if the predetermined number of data is reached when the window is closed.
  • Page 19: Instant Printout

    MEASUREMENT 5.6 Instant printout If a mobile printer is connected, a consecutive printout can be made. The measured data are then instantly printed. The instant printout mode stays active when the measuring window is closed. It must be explicitly deactivated (see chapter 6.2.1).
  • Page 20: Measuring Parameter

    MEASURING PARAMETER 6 Measuring parameter 6.1 Description After power-up, the alphaDUR mini restores the last used measuring parameter. The measuring parameters in detail: M a t e r i a l : The currently selected material calibration. For details about material calibration see chapter 8.
  • Page 21: Managing Sets Of Measuring Parameter

    MEASURING PARAMETER 6.2 Managing sets of measuring parameter In the alphaDUR mini combinations of measuring parameters can be stored under a user-defined name. In this way, the measuring parameters required for specific applications can be called up easily. The following parameter are stored: ...
  • Page 22: Loading Measuring Parameter

    MEASURING PARAMETER 6.2.3 Loading measuring parameter Choose menu item Measuring parameter / Load to read back a set of measuring parameter. Select the measuring parameter set from the list by means of the cursor keys. Then press ENTER to load the parameter. 6.2.4 Deleting a set of measuring parameter Select menu item Measuring parameter / Delete to delete a set of measuring parameter which is no longer needed.
  • Page 23: Hardness Conversion

    Edit / Hardness scale (see chapter 6.2.1) or press the SCALE key in the measuring window. 8 Material calibration The alphaDUR mini has to be calibrated for every material that should be measured. These calibrations can be permanently stored to the device.
  • Page 24 MATERIAL CALIBRATION A factory-provided standard calibration for steel is preset in the alphaDUR mini and cannot be deleted. The calibrations can be grouped to sections. So a two-level hierarchy is established to keep things clear if lots of calibrations shall be saved. E.g.
  • Page 25 After the calibration parameter have been set, the calibration can be continued by cursor right ►. Now, the measurements for calibration must be taken. The alphaDUR mini sounds a beep at the end of each measurement. The probe should be lowered evenly and steadily.
  • Page 26: Memory Functions

    Now, the new calibration can be selected in Measuring parameter / Edit / Material. 9 Memory functions The alphaDUR mini can store up to about 100.000 readings. These readings are organized in measurement series (groups). When a new series is created, name for it must be entered. By these names the series can be selected for display and printout.
  • Page 27: Creating A New Series

    USB flash drive under the menu item Memory functions / Copy to USB flash drive. Thus it is possible, to call up series, without connecting the alphaDUR mini to the computer. (see chapter 17 Option: Copying of series to an USB flash...
  • Page 28: System Settings

    SYSTEM SETTINGS 10 System settings 10.1 Language Select the menu item System / Language to set the language. Select the desired language by means of the cursor keys. The chosen language will become active when you close the dialog by pressing ENTER. 10.2 Time Select the menu item System / Time to set the current time.
  • Page 29: Key Mat

    SYSTEM SETTINGS Select: Open dialogue If the new hardness scale should be selected in a dialogue. Next scale If the hardness scale should be set automatically to the next valid scale. 10.4.3 Key MAT In the menu item System / Configuration / key MAT, it is adjusted whether a dialogue for the selection of the new material should be opened by activating the key MAT in the measurement window, or whether the material should be set automatically to the next valid material...
  • Page 30: Tensile Strength Unit

    Under the menu item System / Unlock options can the additional purchased options be unlocked. The license is delivered on an USB flash drive. After conncecting the flash drive to the alphaDUR mini, the option can be selected and unlocked.
  • Page 31: The Uci Method

    THE UCI METHOD 12 The UCI method The UCI method (Ultrasonic Contact Impedance) is successfully used in hardness testing since many years. A rod is excited into a longitudinal oscillation. At the tip of the rod, a Vickers diamond is placed. This diamond is pressed to the specimen with a discrete test load.
  • Page 32 THE UCI METHOD The advantages of the UCI method are the ease of automation and the very good reproducibility of the hardness readings. The reproducibility of the measurements is better than with optical methods of hardness testing because the total area of contact (proportional to d ) enters into the measurement and not only the diagonal d or a diameter.
  • Page 33: Proper Disposal

    PROPER DISPOSAL 13 Proper disposal Consumers are legally required to dispose of batteries at suitable collection points, vending points or dispatch bays. The crossed-out wheeled bin means that batteries must not be disposed of in the household waste. Pb, Cd and Hg designate substances that exceed the legal limits.
  • Page 34: Español

    PROPER DISPOSAL 13.4 Español Los usuarios están obligados por ley a depositar las pilas viejas en un punto de recogida adecuado /punto de venta/centro de envío. El contenedor de basura tachado significa: la pilas no deben desecharse en la basura doméstica. Pb, Cd y Hg designan sustancias que se encuentran por encima de los valores establecidos por ley.
  • Page 35: Technical Data

    Part 1). Measurement is done with test load applied. Indenter Diamond, Vickers pyramid angle 136°. Preferably metals, to which the alphaDUR mini can Test materials be calibrated with standard hardness test blocks. Tests on ceramics, glass and plastics are also...
  • Page 36 TECHNICAL DATA Weights Device 320 g Probe 190 g...
  • Page 37: Option: Copying Of Series To An Usb Flash Drive

    The license for these options is bound to the alphaDUR mini and is delivered on a USB flash drive. Connect it to the USB interface of the alphaDUR mini. In the menu item System / Unlock options the option can be unlocked then.
  • Page 38: Appendix 1: License Information

    License (GPL) or under the GNU Lesser General Public License (LGPL). The free software source code can at least for a period of 3 years be requested from BAQ GmbH. However, please be noted that we cannot provide guarantee with the source code, and there is also no technical support for the source code from us.
  • Page 39 APPENDIX 1: LICENSE INFORMATION DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED...
  • Page 40 BAQ GmbH Hermann-Schlichting-Straße 14 38110 Braunschweig Germany Tel: +49 5307 95102 - 0 Fax: +49 5307 95102 - 20...