Replacement Parts And Accessories - Fluke Series II User Manual

Hide thumbs Also See for Series II:
Table of Contents

Advertisement

Electrical (cont.)
Measurement
Accuracy, T1, T2,
or T1-T2
(Model 52)
Temperature
Coefficient
Electromagnetic
Compatibility
Maximum
Differential
Common Mode
Voltage
Temperature
Scale
Shop for Fluke products online at:
J-, K-, T-, and E-type: ±[0.05 % of
o
o
reading + 0.3
C (0.5
F)]
o
o
[below -100
C (-148
F): add 0.15 % of
reading for J-, K-, E-, and N-type; and
0.45 % of reading for T-type]
o
o
0.01 % of reading + 0.03
C per
C
o
o
(0.05
F per
F) outside the specified
+18
o
o
o
F to +82
o
C to 28
C (+64
F) range
o
o
[below -100
C (-148
F): add 0.04 % of
reading for J-, K-, E-, and N-type; and
0.08 % of reading for T-type]
Susceptibility: ±2
o
o
C (±3.6
F) for 80 MHz
to 200 MHz in 1.5 V/m field, for 200 MHz
to 1000 MHz in 3 V/m field.
Emmisions: Commercial Limits per
EN50081-1
1 V (Maximum voltage difference
between T1 and T2)
ITS-90
MyFlukeStore
www.

Replacement Parts and Accessories

Applicable
NIST-175
Standards
Accuracy is specified for ambient temperatures between 18
o
o
o
(64
F) and 28
C (82
F) for a period of 1 year. The above
specifications do not include thermocouple error.
Replacement Parts and Accessories
Accessory
Holster and Flex Stand
Assembly
AA NEDA 15A IEC LR6 batteries
80PK-1 K-Type Bead Thermocouple
CD-ROM
Service Manual
.ca
51 & 52 Series II
o
C
Part Number
1272438
376756
773135
1276106
1276123
15
1.800.561.8187

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

5152

Table of Contents