Requirements; Recommended Meter Sizing; Device Type And Software Requirements - ABB CoriolisMaster FCB100 Product Manual

Hide thumbs Also See for CoriolisMaster FCB100:
Table of Contents

Advertisement

FCB100, FCB400 COMMISSIONING AND PROVING | PG/CORIOLIS/PROVING-EN REV. A
Change from one to two columns

1 Requirements

Recommended meter sizing

The following points should be observed when sizing the Coriolis
meter for proving applications:
At flow velocities of 3 to 55 ft/s (1 to 17 m/s), the
optimum flow signal stability for proving is achieved.
Over wider turndowns, it may be necessary to utilize a
different meter factor for lower rates.
Minimum flow rates with 0.8 specific gravity fluid are
recommended as shown below.
Nominal diameter
DN 15 (½ in)
DN 25 (1 in)
DN 50 (2 in)
DN 80 (3 in)
DN 100 (4 in)
DN 150 (6 in)
The proving measurement time should be sufficient to
get a stable flow, and provide enough pulses to the
counter.
The flowing velocity of 55 ft/s (17 m/s) through the meter
should not be exceeded.
Cavitation or void fraction in the fluid should be avoided.
Noise or vibration at frequencies that affect the sensor
should be eliminated.
Flow and density instability should be avoided.

Device type and software requirements

The following settings are valid for these device types:
Minimum flowrate
4 bbl/h (168 ugal/h)
15 bbl/h (630 ugal/h)
45 bbl/h (1890 ugal/h)
115 bbl/h (4830 ugal/h)
220 bbl/h (9240 ugal/h)
270 bbl/h (11340 ugal/h)
CoriolisMaster FCB130 / FCB150 all firmware versions
CoriolisMaster FCB430 / FCB450 all firmware versions
3