Decagon Devices KD2 Pro Operator's Manual page 53

Thermal properties analyzer
Hide thumbs Also See for KD2 Pro:
Table of Contents

Advertisement

1. Heat or cool the sample (with sensor inserted) in a water
bath.
2. Once the sample temperature has equilibrated to the
desired water bath temperature, turn the water bath off.
3. Allow enough time for the water bath to become abso-
lutely still, and make the measurement.
NOTE: Experimental results from researchers have shown that the
KD2-Pro KS-1 sensor can make accurate measurements in water and
aqueous solutions up to about 50 C. Above this temperature, the viscos-
ity of the water becomes too low and free convection begins to affect the
measurement.
For a more in-depth discussion of measuring thermal proper-
ties of liquids, download the application note or watch the
video on this subject from www.decagon.com.
KD2 Pro Operator's Manual
6. Measuring Liquids
49