7.6.15

Set up 2 - Controller quick adjustment

In this menu, you can make a quick correction to the controller set point:
To call the menu, select:
Ñ
"
> Specialist > Set up 2 > Contr. quick adj."
Function
Controller quick adjustment
pH 7.00
Hold
Quick adjustm.
P aram
off
Function
7.00 pH
set
act.
6.58 pH
Select [®]
Next [E ]
7.6.16

Set up 2 - Topcal

Use this menu item to configure the cleaning and calibration cycles and how cleaning and
calibration are triggered.
Configuration of the programs
The following cleaning and calibration programs are available in Topcal S:
• Clean: Predefined program for cleaning the sensor
• Clean S: Predefined program for cleaning and sterilizing the sensor
• Clean C: Predefined program for cleaning and calibrating the sensor
• Clean CS: Predefined program for cleaning, calibrating and sterilizing the sensor
• User 3: Predefined program with short program times to check the system quickly
• User 1/2: Free program slots without predefined program steps
The predefined programs are used for simple programming. You can, however, configure all the
programs as you wish to adapt them optimally to your needs and requirements.
To configure the programs, select "
Note!
• In the programs, you can use external additional valves as required, e.g. for superheated steam, a
second cleaner, cooling air, organic cleaners etc. You control the additional valves with the "Valve
x open", "Valve x closed" program steps.
Controlling the cleaning and calibration programs
You can choose from the following ways to control cleaning and calibration programs:
• Automatic:
Weekly program that automatically starts the selected cleaning program for every weekday. You
can freely select the programs in question for every weekday.
• Cleaning:
Select the cleaning program that is started in the event of an SCS alarm (see "Set up 2 - Check
systems" section) or as per the configured error messages (see "Set up 1 - Alarm" section).
• Power failure program:
Select the cleaning program that is automatically started after a power or air supply failure or if
communication fails.
• Ext. control:
The cleaning and calibration programs can be started by means of a process control system. The
programs are started by means of a 3-bit signal. Please refer to the table in the "Function overview
of the cleaning and calibration programs" section for the binary encoding of the individual
programs.
Please refer also to the "Connecting the external inputs and outputs to the control unit" section
for the electrical connection of the binary encoding for an external program start.
!
Note!
The appendix contains an example of wiring for the external control of the cleaning programs.
Options
Activation and subsequent entry
• Control point
-1.64 ... 15.64 pH
Ñ
> Set up 2 > Topcal > Config. Topcal > Prog. editor".
Info
Enter the set point for the controller function.