Set Up 2 - Controller Quick Adjustment - Endress+Hauser Topclean S CPC30 Operating Instructions Manual

Automation of ph/redox measurements
Table of Contents

Advertisement

Topclean S
PARAM
CHOICE
(default = bold)
pH 7.00
Hold
Selection
Param
Limit switch 1
Limit switch 2
Limit switch 3
Limit switch 4
Limit switch 5
Edit [¯]
[E ]
Next
Limit switch 1 / 2 / 3 / 4 / 5:
pH 7.00
Hold
Param
Configuration
off
Function:
Assign pH/mV Input 1
On value: 16.00 pH
Off value: 16.00 pH
[E ]
Select [ ¯®]
Next
PARAM
Endress+Hauser
To enter the menu, proceed as follows:
Limit switch 1
Limit switch 2
Limit switch 3
Limit switch 4
Limit switch 5
Function
Assign
On value
Off value
On delay
Off delay
Alarm limit
6.6.14

Set up 2 - Controller quick adjustment

In this menu, you can adjust the controller setpoint.
To enter the menu, proceed as follows:
INFO
Selection
of the limit switch which you wish to configure. There are five
limit switches available.
off
Limit switch configuration:
pH/Redox
Function: Activation of function as limit switch.
pH 16.00
Assignment: Selection of the measured value to which the
(1500 mV/
limit value (pH/redox, temperature) is assigned.
100%/150°C)
On point: Entry of the value at which the limit value function
pH 16.00
is activated.
(1500 mV/
Off point: Entry of the value at which the limit value function
100%/150°C)
is deactivated.
(Settable range: pH –2.00 ... 16.00 /
–1500 mV ... +1500 mV / 0 ... 100% / –50 ... +150°C)
0000 s
Limit switch configuration:
0000 s
On delay: Entry of the switch-on delay (range 0 ... 2000 s).
pH 16.00
Off delay: Entry of the switch-off delay (range 0 ... 2000 s).
(150°C)
Alarm limit: Entry of the value (alarm threshold) at which the
fault signalling contact switches.
6 Commissioning
87