Raytech TR-MARK III Instruction Manual

250v option
Hide thumbs Also See for TR-MARK III:

Advertisement

TR-MARK III - 250V Option
(TRO-250)
Instruction Manual
Raytech GmbH
Oberebenestrasse 11
CH 5620 Bremgarten
Switzerland
90174.1.03 Instruction Manual 250V option TR-MARK III
Page 1 of 7

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TR-MARK III and is the answer not in the manual?

Questions and answers

Summary of Contents for Raytech TR-MARK III

  • Page 1 TR-MARK III - 250V Option (TRO-250) Instruction Manual Raytech GmbH Oberebenestrasse 11 CH 5620 Bremgarten Switzerland 90174.1.03 Instruction Manual 250V option TR-MARK III Page 1 of 7...
  • Page 2 Operation ............................5 Creating a 250V Transformer Profile ....................5 Changing Test Voltage in Main Screen ....................5 Import and export of 250V Transformer Profiles................... 6 Technical Specification ........................7 90174.1.03 Instruction Manual 250V option TR-MARK III Page 2 of 7...
  • Page 3: Safety Precautions

    Do not vary the input power source voltage level (IE...Connected to a variable AC power source). The TR-MARK III auto-senses the input power from the mains plus from 100 to 240 vac 50/60Hz. Varying the input voltage will cause the test voltage to vary and result in a higher or lower test voltage than indicated.
  • Page 4 Because of the danger of introducing additional hazards, do not install substitute parts or perform any unauthorized modification to the instrument. Return the instrument to a Raytech service department for service to ensure proper operation and that safety features are maintained.
  • Page 5 Aditional 250V button 2.2 Changing Test Voltage in Main Screen As with a standard TR-MARK III, just touch the test voltage button, and a selection including 250V will appear. 90174.1.03 Instruction Manual 250V option TR-MARK III Page 5 of 7...
  • Page 6 2.3 Import and export of 250V Transformer Profiles If you are using a firmware version 3.0.97 or higher on a TR-MARK III with the 250V option, profiles with 250V test voltage can be imported and exported as usual. When using an older firmware version, or a common TR-MARK III, the test voltage of a 250V transformer profile will be reduced to an available voltage automatically.
  • Page 7: Technical Specification

    601 … 5000 ±0.10° 5001 ... 10’000 ±0.15° 10’001 ... 32’500 ±0.40° 32’501 ... 40’000 ±0.50° Current Resolution: 0.1 mA Range Accuracy test 250V 0 ... 400mA ±1 mA 90174.1.03 Instruction Manual 250V option TR-MARK III Page 7 of 7...