5. Placement of the FSG200 in a room
General Placement Requirements
1 . The feet of the FSG200 must always be placed on a hard level surface .
2 . Do not place the feet on surfaces that will cause them to wobble or be unstable . Make sure the surfaces are
free of any obstacles that may cause the feet to be unstable .
3 . Never place the feet on wet surfaces .
4 . Assemble the FSG200 so that the Track is directly over the location where the transfer will occur . For beds
this is generally between 900mm - 1100mm from the head of the bed . For wheelchairs, living room chairs,
recliner chairs and similar furniture this is directly above the middle (centre) of the seat .
Drawing showing front view of a common FSG200 System set up in a room . In this drawing arrangement,
a general placement of a bed is shown in relationship to the system . Notice that there is still room for a
wheelchair to be placed beside the bed .
DO NOT GO OUTSIDE OF THIS RANGE OF PLACEMENT OTHERWISE INJURY
MAY OCCUR TO THE OPERATOR OF THE LIFT AND/OR THE INDIVIDUAL
BEING TRANSFERRED, AND/OR THE EASY FIT SYSTEM
After Installation
Make sure all four foot feet are sitting flat on the floor . There should be no gap between the feet and floor .
During the use of FSG200 do not bump or lean against the posts . This might cause the system to become
unstable .
In places where more than one operator will be responsible for using the FSG200 , associated hoist and
slings, it is imperative that all operators be trained in its' proper use .
To maintain optimum function, the FSG200 should be inspected and maintained on a regular basis . The
manufacturer 's warranty is void if persons unauthorised by Prism Medical perform work on the FSG200 .
Please ensure adequate clearance after installation to ensure the power supply for the hoist can be accessed .