Hisense Babysense 7 User Manual page 4

The original non-contact baby breathing monitor

Advertisement

Layout of control unit
See illustra�on 7 for layout of control unit featuring:
• On/Off bu�on.
• Green breathing movement indicator light.
• Acous�c and visual red alarm light.
• Low ba�ery red warning light.
• Yellow shutdown indicator light
Red low ba�ery
warning light
Green breathing
movement
indicator light
Opera�on
Babysense is ready for opera�on and use once the
installa�on is completed. To operate, follow the below
instruc�ons:
• Place your baby in crib/bed.
• Turn control unit on by pressing On / Off bu�on.
• Make sure red control lights switch on briefly and a
DOUBLE click sound is heard.
• When breathing and movement are sensed, the green
light flashes.
• Acous�c and visual red alarm light are ac�vated if
baby's breathing stops or slows significantly.
• If red low-ba�ery warning light flashes, replace all
ba�eries as soon as possible. If in addi�on clicking is
heard, all ba�eries must be replaced immediately.
• Turn off by pressing On / Off bu�on before taking
your baby out of crib / bed. Make sure the Yellow
shutdown indicator light momentarily blinks and no
other lights are seen therea�er to indicate the device
is turned off.
Tes�ng Opera�on
Babysense is a highly sensi�ve device, which monitors
baby's breathing movements through the ma�ress
during sleep. There are two main causes of malfunc�on -
technical malfunc�ons and malfunc�ons due to
environmental factors.
Technical malfunc�on
In order to minimize technical problems, follow
installa�on and opera�on instruc�ons carefully.
User Guide
Due to the high sensi�vity of Babysense, fans, wind, air
condi�oning, and so forth can influence it. Such
environmental effects could be picked up by the sensors
on a similar level to that of breathing movements, thus
preven�ng or delaying alarm ac�va�on.
If the Babysense is concurrently used with other devices,
it is required to conduct the daily performance test
(below) under an environment iden�cal to the actual
situa�on and while all concurrently used devices are
"ON". Under such condi�ons, it is required to verify that
the acous�c and visual alarms of Babysense are ac�vated.
Yellow shutdown
indicator light
Red
alarm light
Hidden On/O
Do not conduct performance test by picking up baby
power bu�on
out of bed and wai�ng for the alarm to sound
• The performance test should be conducted while baby
is not in bed.
• Turn device on and take two steps away from bed.
• Ensure that alarm goes off within 20 seconds (due to
absence of breathing movements).
• The alarm may sound with or without preliminary �cking.
• Check that alarm can be heard in all loca�ons
throughout the home or place of care.
I
llustra�on 7
• In case that alarm does not sound within 20 seconds
refer to troubleshoo�ng on page 4-5.
• Performance test must be conducted daily and every
�me you move the monitor, change bedding or
relocate crib/bed posi�on.
Isolated cases of false alarms are possible. Check if
caused by faulty connec�on of the sensor pad(s) to the
control unit, connec�on of only one sensor pad when
baby can move off the immediate area or failure of the
sensor pads to be directly against the ma�ress.
In case of false alarms, kindly refer to troubleshoo�ng
guide on page 4-5. If the problem cannot be solved,
please contact manufacturer's service personnel.
Act immediately according to the following:
1. Approach the bed and check that baby is breathing.
Note pallor (color of skin), muscle tone and movement.
2. If baby does not seem normal, pinch baby's heel and
check that there is no cause for choking, such as an
object lodged in baby's throat.
When breathing resumes, consult a doctor.
3. If breathing does not resume, apply first aid and seek
medical assistance immediately.
3
3
Environmentally caused malfunc�on
Conduct this performance test daily
False alarms
In the event of alarm