Repair And Warranty Orders; Spare Parts - BeaconMedaes Lifeline Camel Installation, Operation And Maintenance Manual

Lpv liquid ring medical vacuum system
Table of Contents

Advertisement

"Camel" Liquid Ring Medical Vacuum
15. Install rotor washer (110-3) on shaft.
16. Perform step j.12, Section B.6.4.
17. Perform step j.l3, Section B.6.4, and fill lathe center in end of shaft with Loctite 262.
18. Perform steps f.l4, f.15 and f.l6, Section B.6.4.
19. Perform steps I and a, Section B.6.4.
20. Perform steps e and g.
21. Install new flange gaskets (138-3 and 139-3) and secure inlet and discharge flanges (138 and
139) to pump.
i. Install drain plug (22) in head (103). This completes re-assembly of pump. Pump must be realigned in
accordance with instructions in Bulletin No.642, Installation Instructions, Nash Vacuum Pumps and
Compressors, if pump was removed from base for disassembly. Contact your BeaconMedæs
representative for more information. Drain and flush as specified in Section B.3.3 and start unit as
specified in Section B.3.4 and Section B.3.5 to place pump back in service.
MAKE CERTAIN THAT LIQUID COMPRESSANT (SEAL LIQUID) SUPPLY IS
TURNED ON BEFORE STARTING PUMP MOTOR. SHAFT SEAL WILL BE
B.7

Repair and Warranty Orders

If a warranty is claimed, the pump is to be shipped to our company in a closed package. It is to be emptied and
cleaned before packaging. For safety reasons, if the pump has been working with harmful or dangerous liquids it is
to be washed with appropriate materials.
B.8

Spare Parts

When ordering spare parts please refer the part name, item number, the pump model, and also the nameplate
references and the serial number. Some parts are standard and therefore, generally available from BeaconMedæs
stock.
CAUTION
DAMAGED IF PUMP IS RUN DRY.
B-15
Medical Systems