Safety; Operational; Service - BeaconMedaes Lifeline Camel Installation, Operation And Maintenance Manual

Lpv liquid ring medical vacuum system
Table of Contents

Advertisement

"Camel" Liquid Ring Medical Vacuum
B.1

Safety

B.1.1

Operational

While the pump is running the following safety pre-cautions should be adhered to:
- Avoid the suction and discharge connections.
- Do not touch the casing and the cover while the pump is working and conveying hot fluids.
- Do not go close to the pump while it is working with a special or toxic fluid and the mechanical seal is
leaking.
- Do not allow the pump to run for long periods of time, if it is creating loud and prolonged noises.
- Check the safety system periodically.
B.1.2

Service

Prior to any repair service being performed, the following safety precautions should be adhered to:
- Stop the pump.
- Bleed air into/out of the piping so that the pump internal pressure is the same as atmospheric pressure.
- Be sure that the current supply is off, the circuit breaker is open, locked, and tagged out.
- Empty the pump of service liquid.
- If the pump has been operated with a harmful liquid, careful washing with an appropriate liquid will be
necessary.