Inlet Gas Or Vapor; Service Liquid - BeaconMedaes Lifeline Camel Installation, Operation And Maintenance Manual

Lpv liquid ring medical vacuum system
Table of Contents

Advertisement

"Camel" Liquid Ring Medical Vacuum
A.3.2

Inlet gas or vapor

The inlet gases or the vapor/gas mixture should not contain solid particles. However, the pump can handle small
quantities of liquids or powders. If the temperature of the inlet gas or vapor is higher then160°F, the service liquid
flow rate should be increased by up to 50% over the flows specified in Table A-2.
A.3.3 Service Liquid
While the pump is operating, it must continuously be fed with clean service liquid. Solid particles in the service
liquid, such as sand or mud will reduce the life of the pump. If solids are present in the service liquid, it is
recommended that a strainer be added to the seal piping plan. Service liquid temperature should be maintained at all
times to assure proper operation. Higher than recommended service liquid temperatures will result in reduced pump
capacity. In certain high inlet temperature cases, it is recommended to use a pre-condenser in order to improve
pump cooling.
Table A-1 Vacuum Operating Limits
Maximum Rotation
and Speed
Maximum Vacuum
Maximum Discharge
Pressure
Maximum Inlet Gas
Temperature
Maximum Service
Liquid Temperature
Table A-2 Seal Water Flow Rates
System Size - Pump
3 HP
5HP, 7½ HP
10 HP
3500 RPM (EC 50)
1800 RPM (EC 90/125/150/250)
29" HgVac/1.0 HgAbs
(980 mbar/33 mbar Abs)
0.5 PSIG (0.03 Barg)
212 ° F (100 ° C)
160 ° F (70 ° C)
Flow Rate - GPM*
0.50 GPM
0.75 GPM
1.0 GPM
*Per vacuum pump
A-4