Instruction Manual ® FLOW series Digital Mass Flow / Pressure Controllers for gases Doc. no.: 9.17.045V Date: 26-11-2018 ATTENTION Please read this Instruction Manual carefully before installing and operating the instrument. Not following the guidelines could result in personal injury and/or damage to the equipment.
Page 2
Disclaimer This document has been reviewed and is believed to be accurate. Bronkhorst High-Tech B.V. does not assume liability for errors, inaccuracies or absence of information. The material in this document merely serves information and illustration purposes;...
Page 3
Bronkhorst® Warranty Bronkhorst® products are warranted against defects in material and workmanship for a period of three years from the date of shipment, provided they are used in accordance with the ordering specifications and not subject to abuse or physical damage.
Page 4
Bronkhorst® Instruction Manual IQ+FLOW® Series Digital Mass Flow / Pressure Controllers for Gases 9.17.045V...
· Dimensional drawing IQF/IQP (document no. 7.05.870) · Dimensional drawing IQFD/IQPD (document no. 7.05.871) · Dimensional drawing IQM (document no. 7.05.760) These documents can be downloaded from www.bronkhorst.com/downloads and are available by request at our local sales & service representatives 9.17.045V...
To enable us to help you quickly and effectively, make sure to have the serial number ready whenever seeking contact with Bronkhorst about a specific item. The serial number (SN) is the key to the original purchase order.
However, compliance with the EMC requirements is not possible without the use of proper cables and connector/gland assemblies. Bronkhorst recommends the use of their standard cables. These cables have the right connectors and if loose ends are used, these are marked to help prevent wrong connection. When using other cables, cable wire diameters should be sufficient to carry the supply current, and voltage loss must be kept as low as possible.
Digital operation over RS232 can be established when using the following setup. Using a RS232 cable or a USB- RS232 converter with a PC will allow you to use (free) Bronkhorst® software for Windows, such as FlowDDE and FlowPlot. See also section 3.4. Digital RS485 operation With digital operation over RS485 a bus-system with multiple instruments can be set up.
2.11 Calibration Each IQ+FLOW® instrument is factory calibrated. Bronkhorst certifies that all instruments meet the rated accuracy. Calibration is performed using measurement standards traceable to the standards of the Dutch Metrology Institute (VSL). Calibration certificates are included in the shipment. When operated properly (clean gas, no pressure shocks, no vibrations, no thermal shocks, etc.), regular maintenance is not required.
Bronkhorst® 2.13 Model key 2.13.1 Model key IQF / IQP 2.13.2 Model key IQM * Each channel of an instrument will be provided an additional letter after the serial number (e.g. M########AA). Instruction Manual IQ+FLOW® Series Digital Mass Flow / Pressure Controllers for Gases...
For analog operation either an RJ-45 loose-end cable or an RJ-45 to 9-pin sub-D converter may be used to connect the required signals. When using a Bronkhorst® readout unit use only an RJ- Refer to the “Hook-up diagram IQ FLOW®” or use an 45 cable (7.03.236) in combination with the RJ-45 to 9-...
Bronkhorst® Manual interface: micro-switch, LEDs and rotary switches In this section the manual instrument interfaces are described: Micro-switch · Indication LEDs · Rotary switch · 3.3.1 Micro-switch operation (single-channel versions only) By means of manual operation of the micro push-button switch some important actions for the instrument can be selected or started.
Page 15
Restore factory settings flash All parameter settings (except field bus settings) will be restored to 0,2 sec on, situation of final test at Bronkhorst® production. 0,2 sec off Normal 8 – 12 sec. For FLOW-BUS only: install a free node-address on FLOW-BUS.
Bronkhorst® 3.3.2 LED functions The LEDs on top of the instrument can also be used for manual operation of some options. The green LED will indicate in what mode the instrument is active. The red LED will indicate error/warning situations.
3.4.2 Dynamic Data Exchange (DDE) RS232 communication can be used for operating the instrument using the Bronkhorst® FlowDDE server application. Dynamic Data Exchange (DDE) provides the user a basic level of interprocess communication between Windows applications. FlowDDE is a DDE server application. Together with a client-application, either self-made or with a SCADA- program from third parties, it is possible to create an easy way of data exchange between the flow controller and a Windows application.
DDE are for example MS-Office, LabVIEW, InTouch and Wizcon. Bronkhorst® software applications 'FlowView' (left) and 'FlowPlot' (right) FlowDDE and other Bronkhorst® applications are available on the support CD or can be downloaded from the product pages on the Bronkhorst website: www.bronkhorst.com/products 3.4.5...
Bronkhorst® 3.5.1 Hook-up These illustrations show examples of IQ FLOW® instruments in a RS485 bus-system. When the power consumption is more than 15W, two separate power supplies are required due to the maximum power supply of 15 W for one PiPS.
Software When using a pc to communicate with IQ FLOW® instruments only the FLOW-BUS protocol is supported by Bronkhorst® software. When using Modbus operation, software from third parties, such as LabVIEW, ModScan or a Modbus PLC must be used to serve as Modbus master.
Most instrument parameters can only be accessed with digital communication. For each communication protocol the instrument parameters are accessed differently. When using Bronkhorst® software programs FlowView or FlowPlot, easy access is provided to the mostly used parameters by menu interfaces. When using other communication methods the...
This parameter consists of a maximum 20-byte string with instrument serial number for identification, for example: 'M1111111A'. BHT Model Number Type Access Range FlowDDE FLOW-BUS Modbus Unsigned char[23] 113/2 0xF110…0xF116/61712…61718 ® This parameter shows the Bronkhorst instrument model type information. Instruction Manual IQ+FLOW® Series Digital Mass Flow / Pressure Controllers for Gases 9.17.045V...
Bronkhorst® Advanced operation Reading and changing instrument parameters 4.1.1 Special parameters All parameters described in this chapter affect the behavior of the IQ FLOW®. Please be aware that wrong settings can disorder the output. To avoid unintentional changes, some parameters are locked (shown by the symbol). To unlock parameters set parameter 'Init Reset' to 'Unlocked'.
Bronkhorst® Value Mode Instrument action Setpoint Master Slave source source factor RS232 Controlling RS232 Valve steering Setpoint is redirected to Valve Out with controller idle Analog valve Analog input is redirected to Valve Out with steering the controller idle Valve safe state After power-up, the control mode is set to 'Analog input' or 'BUS/RS232', depending on the customer's default setting for analog or digital operation.
The controlling algorithm for the valve handled by the micro-controller consists of several parameters which can be set. Although many parameters could be accessed, Bronkhorst advises not to change these parameters because during manufacturing the controller is optimized. Changing of controller settings should be performed by or under supervision from trained service personnel only.
Bronkhorst® Controller Speed (Kspeed) Type Access Range FlowDDE FLOW-BUS Modbus Float RW 0…3.4E+38 114/30 0xF2F0…0xF2F1/62193…62194 This parameter is the controller speed factor. PID-Kp is multiplied by this factor. Ti (PID-Ti) Type Access Range FlowDDE FLOW-BUS Modbus Float RW ...
Bronkhorst® Dynamic Display Factor Type Access Range FlowDDE FLOW-BUS Modbus Float RW 0…1.0 117/1 0xF508…0xF509/62729…62730 This value should not be changed. Static Display Factor Type Access Range FlowDDE FLOW-BUS Modbus Float RW 0…1.0 117/2 0xF510…0xF511/62737…62738 This value should not be changed.
Bronkhorst® 4.2.2 Changing default control mode Instruments are delivered with either analog or digital signal as default, depending on customer's requirement. After every (power-up) reset the instrument will return to its default control mode. The default control mode can be changed with the...
Bronkhorst® 4.2.4 Setting digital output (multi-channel versions only) The IQ FLOW® multi-channel pc-board is equipped with three digital outputs. The digital outputs can be used for driving shut-off valves (for instance). The digital outputs can be read or written via the parameter 'IO Switch Status'. The parameters can be set as indicated in the value table below.
Page 30
Bronkhorst® Value zero will be indicated by a period of 1 sec off (0.5 sec off + 0,5 sec off). When value zero is wanted, press switch shortly and release it again within 1 sec. Before each action of flash-counting, the LEDs to be used for counting will flash in a high frequency.
Bronkhorst® Troubleshooting and service General For a correct analysis of the proper operation of an instrument it is recommended to remove the unit from the process line and check it without applying fluid supply pressure. In case the unit is dirty or clogged, this can be ascertained immediately by loosening the fittings and inspecting visually.
To enable us to help you quickly and effectively, make sure to have the serial number ready whenever seeking contact with Bronkhorst about a specific item. The serial number (SN) is the key to the original purchase order and can be found on the product.
Within the scope of the European Union, manufacturers of electrical and electronic equipment (EEE) are bound to comply with the WEEE Directive (Waste Electrical and Electronic Equipment). As a consequence, Bronkhorst is obligated to offer its customers in the EU the possibility to return EEE for disposal once it has reached the end of its lifetime, and take all necessary steps to dismantle it properly and recycle or re-use its components whenever possible.