To vent the MSD without the ChemStation

To vent the MSD without the ChemStation

If the MSD ChemStation is functioning, use the procedure on page 52. This proce-
dure should only be used if it is absolutely necessary.
1 If your system is equipped with a gauge controller, switch off the triode
gauge tube and gauge controller.
2 From the GC keypad, turn off the GC zone controlling the GC/MSD
interface heater and set the GC oven to 30°C (see page 49).
3 Disconnect the
This turns off the MS heaters and the diffusion pump heater or turbo pump, but
the cooling fans and foreline pump stay on.
4 After 45 minutes, turn off the MSD power switch.
5 Unplug the MSD power cord.
: $ 5 1 , 1 *
,I \RX DUH XVLQJ K\GURJHQ DV D FDUULHU JDV WKH FDUULHU JDV IORZ PXVW EH RII EHIRUH WXUQLQJ
RII WKH 06' SRZHU ,I WKH IRUHOLQH SXPS LV RII K\GURJHQ ZLOO DFFXPXODWH LQ WKH 06' DQG DQ
H[SORVLRQ PD\ RFFXU 5HDG WKH +\GURJHQ &DUULHU *DV 6DIHW\ *XLGH  EHIRUH
RSHUDWLQJ WKH 06' ZLWK K\GURJHQ FDUULHU JDV
& $ 8 7 , 2 1
Be sure the GC oven and the GC/MSD interface are cool before turning off carrier gas flow.
6 Reconnect the
7 Remove the upper MSD cover (page 50).
8 Turn the vent valve knob counterclockwise to admit air into the vacuum
manifold.
Do not remove the knob. Be sure to retighten the knob before pumping down.
See Also
To open the vacuum manifold, page 54
64
Operating the MSD
HIVAC POWER
HIVAC POWER
cable from the back panel.
cable.