To Replace The Fan For The High Vacuum Pump - HP 5973 Hardware Manual

Mass selective detector
Hide thumbs Also See for 5973:
Table of Contents

Advertisement

4

To replace the fan for the high vacuum pump

To replace the fan for the high vacuum pump
Materials needed:
Fan (3160-1037)
Screwdriver, TORX T-15 (8710-1622)
1 Vent the MSD (page 52).
2 Remove the upper and lower MSD covers (page 50).
3 Disconnect the fan wiring from the connector on the MSD frame.
4 Remove the 4 fan screws and remove the fan.
Keep the 4 screws.
: $ 5 1 , 1 *
'R QRW WRXFK WKH KLJK YDFXXP SXPS 7KH KLJK YDFXXP SXPSV HVSHFLDOO\ WKH GLIIXVLRQ SXPS
RSHUDWH DW GDQJHURXVO\ KLJK WHPSHUDWXUHV DQG FRXOG VWLOO EH KRW HQRXJK WR EXUQ \RX
5 Disconnect the fan wiring and safety grill from the old fan.
The fan wiring ends in a small connector on the back of the fan
6 Connect the fan wiring and safety grill to the new fan.
7 Install the new fan and reinstall the 4 screws.
: $ 5 1 , 1 *
0DNH VXUH WKH VDIHW\ JULOO WKDW VKLHOGV WKH IDQ EODGHV LV LQ SODFH
8 Connect the fan wiring to the fan connector on the MSD frame.
9 Reinstall the MSD covers.
10 Pump down the MSD (page 58).
134