Setup-Menu Functions - Radcal Accu Series User Manual

Radiation measurement systems
Table of Contents

Advertisement

Setup–Menu Functions
Pressing select at the main menu display
enters Setup. The following information and options are provided in this mode::
Serial No. and firmware Version
!
This manual applies to version 6.
Pressure and temperature
!
Values used for ion-chamber compensation. Prefixed with Std if measured not selected.
SELECT restores setup defaults.
!
Restores all items in Setup to default values. Default values are shown as Default in the following
descriptions.
See detailed instructions on page 8
Mode SELECT
!
Parameter
Pressure
Temperature
Len/area suffix
1
High sensitivity
Radiation units
Auto shutdown
Time units
Mammo anode/filter
DAC output
1
The Accu-Pro measures the background zero when you select a dose mode. If high sensitivity is set to
off, the system will make a quick coarse zero measurement. Some fluctuations will be observed in dose
rate mode and dose accumulate may exhibit drifting. If high sensitivity is set to on, the system will make
a fine zero measurement which takes 3 to 5 seconds. The dose rate will fluctuate much less and the
dose accumulate drift will be greatly reduced. It is recommended that high sensitivity mode is enabled for
most dose measurements.
Function SELECT
!
Add/Remove functions from the main menu. See Page 17 for a list of functions
kV disp(lay) SELECT
!
See table on page 12
Intrinsic filt(ration) SELECT
!
Selects the filtration used to compute kV in the diagnostic range. Auto enables automatic calculation
based on beam hardness.
Auto, Cu (32 steps from 0 to 0.695 mm), Al (32 steps from 0 to 23 mm).
Change or SELECT
Choices
Measured/Std 101.3 kPa
Measured/Std 22 deg C
On(*cm or *cm2)/Off
On/Off
Roentgens/Grays/Coulombs
On(30 min)/Off
Seconds/Minutes/Hours
Mo/Mo, W/Ag, W/Rh
kV/mA/Rel. intensity/dose rate
setup
Accu-Pro
19
Accu-Dose Accu-kV
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X