REFERENCE

Display

The 2-line 16-character LCD displays menu selections and measurement results.
Characters positioned at the left side of the display have special meaning:
> in Setup indicates a parameter choice. Press select to chose the item.
For dose measurements:
? indicates dose-sensor signal was outside calibrated range .
> indicates high-sensitivity mode measurement.
[ indicates low -sensitivity mode measurement.
] indicates cumulative (zero-corrected) dose is negative.
For kV measurements:
W means Wait. The kV system is unable to process another exposure while W is displayed.
A indicates automatic filtration compensation is enabled.
D indicates default filtration will be used:
M indicates a user-selected filter (other than 2.5 mm Al).
T indicates a Tungsten/Rhodium anode/filter
S indicates a Tungsten/Silver anode/filter
In the lower right corner of the display:
C indicates a dose sensor (ion chamber or diode)
K indicates dose from the Accu-kV diode
The kV display may be configured in Setup. The following table shows the selections for each of the four
locations on the display. Default values are underlined.
Location
Upper Left
kVp
Upper
mAs
Right
Lower
conv dose
Left
Lower*
width ms
Right
* Readout not present if Lower Left is Conv Dose in kV
2.5 mm Aluminum for the diagnostic sensor.
30μm Molybdenum for the Mammography sensor
Measurement
kV practical
width
pulses
width
pulses
kV diode
dose rate
12
kV
reduced
width
ms
width
ms
rel
dose rate
kV avg
mA
mA
pulses