The following is a brief description of the parameters and the available range of settings:
Parameter
SpO
Output
2
Auto Off Timer
Backlight Timed
Backlight Intensity
Battery Life
Contrast Adjust
Power up with
Auto Step Time
Software
Description
Sets the ability to drive an external
SpO
module (MSP-2100)
2
The elapsed time after which the unit
will automatically power down. This
timer
is
reset
depression.
(Setting the value to 0 eliminates this
feature.)
Off – Always off
1-30 sec – The elapsed time after
which the backlight will automatically
turn off.
Manual – The backlight will be
manually controlled by backlight key.
Sets the intensity of the backlight.
(Note:
Lower intensities extend
battery life.)
Displays current life of the batteries.
At 5%, a warning screen will appear.
At 0%, the unit will power down
automatically.
Sets the contrast of the display
screen.
Selects the values that will be used
when the unit is first turned on. It is
also used to Set the Custom
Defaults, if used.
Settings).
Sets the interval that is used with the
Auto increment features in BPM and
Performance.
Displays current software program.
30
by
each
key
(See Power Up
Range
Enabled/Disabled
0-30 min
Off, 1-30 sec, Manual
0-100%
5-100%
(Read Only)
0-20
Default/Last/Custom/
Set Custom Defaults
1 to 60 sec
(Read Only)