The following is a brief description of how the PS-2105 simulates the available
Pacemaker Waveforms:
Abbreviation
Atrial Pacer
Asynchronous
Non-Capture
Non-Function
Demand –
Occasional
Demand –
Frequent
AV –
Sequential
PACEMAKER
Waveform
Atrial Pacemaker Wave
Asynchronous Pacemaker Wave
Ventricular Pacemaker Wave with
Periodic Non-Response
Ventricular Pacemaker Wave with
no Heart Response
Demand Pacemaker Wave with
Occasional Sinus Beats
Demand Pacemaker Wave with
Frequent Sinus Beats
AV-Sequential Pacemaker Wave
80 BPM with Pacer Pulse at the start
of each P wave
75 BPM with Pacer Pulse at the start
of each QRS wave and no P wave
75 BPM Ventricular Paced beats
with every tenth beat not responding
75 BPM Ventricular Paced beats
with no heart response
20 NSR beats followed by 20
Ventricular Paced beats
40 NSR beats followed by 40
Ventricular Paced beats
75 BPM with Pacer Pulse at the start
of both the P and QRS waves