LaserLiner Cubus Manual page 22

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 13
Preparing the calibration check
It is possible for you to check the calibration of the laser. To do this, position
the device midway between 2 walls, which must be at least 5 metres apart.
Switch the device on. The best calibration results are achieved if the device
is mounted on a tripod. IMPORTANT: The automatic sensor must be active
(auto/slope LED is off).
1. Mark point A1 on the wall.
2. Turn the device through 180° and mark point A2. You now have a
horizontal reference between points A1 and A2.
1.
A1
2.
A1
Performing the calibration check
3. Position the device as near as possible to the wall at the height of point
A1. Now adjust the device in the X axis.
4. Turn the device through 180° and mark point A3. The difference between
points A2 and A3 is the tolerance for the X axis.
5. To check the Y and Z axis, repeat steps 3 and 4.
3.
A1
4.
If points A2 and A3 are more than 0.15 mm / m apart on either
!
the X or Y axis, the device is in need of adjustment. Contact your
authorised dealer or else the UMAREX-LASERLINER Service Department.
22
EN
A2
A2
A2
A3
A2
A3

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cubus g

Table of Contents