P6.0: Resetting to factory default settings
8.8
This program lets you reset all parameters to factory default settings. This clears all
calibration data, memory, and any other setup functions you might have changed.
However, the following settings will remain as you have set them:
• temperature unit of measure (°C or °F)
• The temperature offset calibration value
• the LCD contrast setting
• the clock setting.
From measurement mode
1.
Press Setup key to enter Set Up mode.
2.
Press the
and
subgroups until you view parameter
"rSt" in the upper display.
See figure
A
3.
Press the ENTER key.
B
See figure
4.
Press the
and
between NO and YES.
• NO retains current settings
• YES resets to factory default settings
5.
Press the ENTER key to confirm
selection and to return to measurement
mode.
Notes
See page 47 for a table of factory default
settings.
38
keys to scroll through
keys to toggle
A
B