15. Appendix 4: Meter factory default settings

Resetting the meter to factory default settings clears all calibration data and most
other setup functions you might have changed. The following settings will remain as
you have set them:
• temperature unit of measure (°C or °F)
• The temperature offset calibration value
NOTE: Conductivity/TDS and pH data are cleared separately from each other.
• To clear pH data, see Sub group P4.0 on page 32
• To clear conductivity/TDS data, see Sub group P9.0 on page 41
Type Parameter
Default
pH parameters
P1.1
Viewing pH
– – –
P1.2
calibration data
– – –
P1.3
– – –
P1.4
– – –
P1.5
– – –
P2.1
View electrode offset
0.0 mV
P2.2
View electrode slope
100.0%
P3.1
Ready indicator
Ready on
P3.2
# pH cal. points
3
P3.3
USA or NIST buffers
USA
P3.4
°F or °C
no default
P4.0
pH factory default
no
Conductivity/TDS parameters
P5.1
Viewing con./TDS
– – –
P5.2
calibration data
– – –
P5.3
– – –
P5.4
– – –
P5.5
– – –
P6.1
Viewing con./TDS
1.000
P6.2
calibration data
1.000
P6.3
1.000
P6.4
1.000
P6.5
1.000
P7.1
Ready indicator
Ready on
P7.2
°F or °C
no default
P7.3
ATC on or off
ATC on
P7.4
Set TDS factor
0.5
P8.1
Temp. coefficient
2.1% per °C
P8.2
Normalization temp.
25°C
P9.0
factory default
no
50
Remarks
No calibration data for pH 1.68
No calibration data for pH 4.01
No calibration data for pH 7.00 USA (6.86 NIST)
No calibration data for pH 10.01 USA (9.18 NIST)
No calibration data for pH 12.45
No offset adjustment
No slope adjustment
Ready indicator on; auto endpoint off
3 pH calibration points available (1-5 point range)
USA buffers: pH 1.68, 4.01, 7.00, 10.01, 12.45
°F or °C remains as selected
Retains your current settings
No cal. data for R1 (0.00-19.99 µS/0.00-9.99 ppm)
No cal. data for R2 (0.0-199.9 µS/10.0-99.9 ppm)
No cal. data for R3 ( 0-1999 µS/100-999 ppm)
No cal. data for R4 (0.00-19.99 mS/1.00-9.99 ppt)
No cal. data for R5 (0.0-199.9 mS/10.0-199.9 ppt)
No offset for R1 (0.00-19.99 µS/0.00-9.99 ppm)
No offset for R2 (0.0-199.9 µS/10.0-99.9 ppm)
No offset for R3 ( 0-1999 µS/100-999 ppm)
No offset for R4 (0.00-19.99 mS/1.00-9.99 ppt)
No offset for R5 (0.0-199.9 mS/10.0-199.9 ppt)
Ready indicator on; auto endpoint off
°F or °C remains as selected
Automatic Temperature Compensation on
Adjustable from 0.4 to 1.0
Adjustable from 0 to 10% per °C
Adjustable from 15 to 30°C / 59 to 86°F
Retains your current settings