Program Cof: Unit Configuration; Ready Indicator And Auto Endpoint Function - Accumet AP74 Operating Instructions Manual

Portable waterproof dissolved oxygen meter
Table of Contents

Advertisement

COF: Unit configuration
8.7
Unit configuration mode lets you select the following parameters:
A. Ready indicator and auto endpoint function
B. mg/l or ppm units (available from mg/l or ppm mode only)
C. Temperature in °C or °F

READY indicator and auto endpoint function

The first program lets you select:
• "READY indicator on" to indicate when the reading is stable.
• "READY indicator off" for faster meter response.
• Auto endpoint function on. Select auto endpoint on to "hold" the reading when
it is stable for more than 5 seconds. The display automatically freezes, and the
HOLD indicator appears on the left side of the display. Press the HOLD key to
release the display and access other functions.
From measurement mode
1.
Press Setup key to enter Set Up mode.
2.
Press the
and
subgroups until you view parameter COF
in the upper display.
See figure
A
3.
Press the ENTER key to select
parameter rdY (Ready).
See figure
B
4.
Press the
and
configuration you require.
•ON switches the READY indicator on.
•OFF switches the READY indicator off.
• ON and HOLD together switches the
auto endpoint feature on.
5.
Press the ENTER key to confirm selection
and to proceed to:
• in % mode: step 4 on page 34.
• in mg/l (ppm) mode: step 3 on page 33.
Or, press the CAL/MEAS key twice to
return to measurement mode.
Notes
Meter default is set for Ready indicator on,
and auto endpoint function off.
32
keys to scroll through
keys to select the
A
B