HP 6811B User Manual page 4

Hp dual power analyzer
Table of Contents

Advertisement

Front Panel Menus - Additions
The Power Analyzer input uses the same front panel menu commands that are used to measure the Main
Output of the ac source. These are documented in the ac source User's Guide. The following menu
commands have been added to let you access the Power Analyzer input, control the output frequency
modulation, and calibrate the Power Analyzer input. In most cases you will need to press the
Function key multiple times to access the menu item. Press the
Phase
Select
ô
Freq
ô
ô
ô
Input
ô
Shift
Cal
ô
ô
With no signal connected to the Power Analyzer input, the φ2 front panel display may
NOTE:
return a very high frequency value. This is normal, since there is no external frequency
for the measurement circuit to detect.
SCPI Programming Commands - Additions
The Power Analyzer input uses the same measurement commands that are used to measure the Main
Output of the ac source. These are documented in the ac source User's Guide. The following SCPI
commands have been added to let you access the Power Analyzer input, control the output frequency
modulation, and calibrate the Power Analyzer input. Additional measurement functions let you measure
transfer time, peak voltage, and phase locked loop performance when evaluating devices such as
uninterruptible power supplies.
CALibrate
:CURRent
:EXTernal
:VOLTage
:EXTernal
:SHUNt
FM
:STATe <bool>
:DEViation <value>
:FREQuency <value>
4
φ1
Displays the Main Output measurement when φ1 is selected
115V 60 Hz
φ2
Displays the Power Analyzer measurement when φ2 is selected
120V 60 Hz
FM OFF
Enable or disable output frequency modulation (OFF or ON)
FM:FREQ 50
Specifies the modulating frequency of the output
FM:DEV 10
Specifies the peak deviation of the output frequency modulation
FREQ:SRC VOLT
Specifies the frequency input source (VOLT or CURR)
CAL:CURR: EXT
Begin external current calibration
CAL:VOLT: EXT
Begin external voltage calibration
SHUNT
Enter a value for the external shunt in ohms
MEASure | FETCh
SENSe
°
Entry key to access the parameter.
:VOLTage
:RANKed? <percentile>
:TVOLt? <level><occurrence>
:ABSolute? <level><start_time><min_pulse_width>
:NSELect 1 | 2
:FREQuency:SOURce VOLT | CURR
ô

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

6812b6813b

Table of Contents