Advertisement

5
Running a Method or a Sequence from the Keypad

Running Methods from the Keypad

To manually inject a sample with a syringe and start a run

To run a method to process a single ALS sample

To abort a method

50
Prepare the sample syringe for injection.
1
Load the desired method. (See
2
Press [Prep Run].
3
Wait for STATUS Ready for Injection to be displayed.
4
5
Insert the syringe needle through the septum and all the
way into the inlet.
Simultaneously depress the syringe plunger to inject the
6
sample and press [Start].
Prepare the sample for injection.
1
Load the sample vial into the assigned location in the
2
ALS tray or turret.
Load the desired method. (See
3
Press [Start] on the GC keypad to initiate the ALS syringe
4
cleaning, sample loading, and sample injection method.
After the sample is loaded into the syringe, the sample is
automatically injected when the GC reaches the ready
state.
Press [Stop].
1
2
When you are ready to resume running analyses, load the
appropriate sequence or method. (See
or
"To load a stored
"To load a
"To load a
"To load a method"
sequence".)
method".)
method".)