Dimensions And Process Connections; Electrical Connection - Siemens SITRANS P ZD Series Operating Instructions Manual

Pressure transmitter
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 37
Operating instructions
4.2

Dimensions and process connections

g
Ø 82 (3.23)
s
M
a
Type A
SITRANS P pressure transmitter, ZD series, dimensional drawing, dimensions in mm (inch)
Figure 3
a) Type A
4.3

Electrical connection

Screened cable (diameter 4.5 to 10 mm) with two wires is routed in via the M16x1.5
cable gland. The cable shield is grounded at one location only (preferably in the
connection cabinet). The grounding connector in the transmitter is connected to the
housing.
1
2
U
Power supply
B
R
Load
L
I
Output current
0
Figure 4
Connection diagram
SITRANS P, ZD series
A5E00249308-01
48,5 (1.91)
45 (1.77)
M16x1,5
SW 27
G½A
½-14 NPT
G½A
EN 837-1
1 (+)
I
0
2 (-)
Installation
Ø 82 (3.23)
M16x1,5
G
1
/
A
8
Type B
b) Type B
+
U
B
R
L
57