Guidance and Manufacturer's Declaration ± Electromagnetic Immunity
The NewLife is intended for use in the electromagnetic environment specified below. The customer or the user
of the NewLife should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
a
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the NewLife Family is used
exceeds the applicable RF compliance level above, the NewLife should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating
the NewLife Family.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
b
26 - ENG
PN MN239 B | User Manual
Compliance level
3 Vrms
3 V/m
Electromagnetic environment ±
guidance
Portable and mobile RF communica-
tions equipment should be used no
closer to any part of the NewLife, in-
cluding cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
d
1
2 .
P
d
1
2 .
P
from 80 MHz to 800 MHz
d
P
1
2 .
from 800 MHz to .5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufac-
turer and d is the recommended sepa-
ration distance in meters (m).
Field strengths from fixed RF transmit-
ters, as determined by an electromag-
netic site survey,
should be less than
a
the compliance level in each frequency
range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol: