8

Specifications

EEG sensitivity
Noise
CMRR
Input impedance
Sample rate
CSI and update
EMG
BS%
Digital output
Wireless range
Display size
Alarms
Artifact rejection
Sensor impedance range
Battery
Supply current
Max. battery lifetime - Alkaline
Max. battery lifetime –
Rechargeable
Recharge time
Weight
Dimensions
Classification
Sensors
Cable length
Mounting options
Memory
Environment – Operation
Environment – Transport
and storage
* If transport without sensors: -4–158°F
Cerebral State Monitor - CSM DFU
±400μV
< 2μV
, < 0.4μV RMS, 1–250 Hz
p-p
>140dB
>50Mohm
2000 samples/sec. (14 bits equivalent)
0-100. Filter 6-42 Hz, 1 sec. update
0-100 logarithmic. Filter 75-85 Hz, 1 sec. update
0-100%. Filter 2-42 Hz, 1 sec. update
Wireless to RS232 link or USB (ISM 2.4 GHz)
Up to 10 meters
1.3 x 0.7"
High / Low with user selectable limit
Automatic
0-10kOhm / measurement current 0,01μA
Rechargeable NiMH or 9V Alkaline
(6AM6/IEC:6LR61/ANSI:1604A )
30mA (typical)
30h (stand alone)
18h (transmitting wireless)
See table in section 6.1
4 Hours (CSMX04 only)
(
4.6 ounces
130 g) with battery
4.6x2.7x1.2"
Internal power supply/ Class II, type BF, continuous use
Danmeter Neuro Sensors
77" with 14" split
Velcro strip 1.7x0.9"
Data recording 18 hours
Temperature
50–104°F
Rel. humidity
30–75%
Air pressure
700–1060 hPa
Temperature
50–104°F *
Rel. humidity
30–95%
Air pressure
700–1060 hPa