Connecting CSM Monitor and CSM Link
1.
Turn on the CSM Link.
2.
Power indicator shows a constant light.
3.
Link indicator is off.
4.
Turn on the CSM monitor.
5.
The CSM monitor shows a list of CSM Links found.
6.
Select CSM Link ID (Serial No.) with the Event key (8) and the indicator Link flashes,
or choose skip in order not to establish a connection. (The radio is reactivated by
pressing the Power On key (1) twice or the next time the CSM is switched on.)
7.
Accept the CSM Link by pressing the Set Event key (6).
8.
When link has been established the indicator Link will show a constant light.
Disconnecting CSM Monitor and CSM Link
1.
Press Power On key twice within one second.
2.
Highlight release using Event key (8).
3.
Press Set Event key (6) to disconnect. CSM radio is inactive. (The radio is reactivated
by pressing the Power On key (1) twice or the next time the CSM is switched on.)
Connecting to CSM Link when the CSM radio is inactive
1.
Turn on the CSM Link.
2.
Power indicator shows a constant light.
3.
LINK indicator is off.
4.
Press Power On key twice within one second.
5.
The CSM monitor shows a list of CSM Links found.
6.
Select CSM Link ID (Serial No.) with the Event key (8) and the indicator Link flashes,
or choose skip in order not to establish a connection.
7.
Accept the CSM Link by pressing the Set Event key (6).
8.
When link has been established the indicator Link will show a constant
light.
Note:
If wireless has been active for more than five minutes without any connection to CSM-Link,
the radio is inactivated.