Standards and Regulatory Listing:
CE marked model
Electrical:
Placement:
Tubing:
Time of Operation:
Recommended Storage and Shipping
Temperature:
Environment:
Minimum Operating Time:
1195656-A
IEC/EN60601-1, A1, A2
IEC/EN60601-1-2
IEC/EN61000-3-2
IEC/EN61000-3-3,
ISO8359
MDD 93/42/EEC, Annex I and IX
IRC5PO2AW/IRC5PO2VAW
No extension cords
No closer than 30,5 cm (12 in) from any wall, furniture, draperies, or furniture to assure sufficient
air flow
Avoid deep pile carpets and heaters, radiators or hot air registers
Floor location only
No confined spaces (Example: No closets)
2 m (7 ft) cannula with a maximum 15 m (50 ft) of Crush-Proof Tubing
(DO NOT pinch)
Up to 24 hours per day
-29° C to 65° C (-20° F to 150° F) at 15-95% relative humidity
Smoke and soot-free
No confined spaces (Example: No closets)
30 Minutes
Technical data