Agilent Technologies 85051B User's And Service Manual page 29

7 mm verification kit
Table of Contents

Advertisement

Figure 3-5
Gaging the Airline
9. Torque the short circuit-to-airline connection and the gage-to-airline connection to 136 N-cm, (12 in-lb)
with the 3/4 inch torque wrench. Set the assembly down to let the temperature stabilize for 5 minutes.
10. The reading on the gage should be within the tolerance band of the dimension given (±0.0002 inch);
±0.001 inch for factory uncertainties and ±0.0001 inch for your gage accuracy.
NOTE
The reading on the gage is an estimate of the actual recession of the center conductor. Gage
accuracy is typically ±0.0001 inch.
Disconnect the airline from the gage and short circuit
1. Remove one device from the airline and replace the plastic cap on the airline to prevent the center
conductor from sliding out of the outer conductor.
2. Remove the device from the other end of the airline. If you won't be using the airline again immediately,
slide the center conductor out of the outer conductor and store the center conductor in the plastic case
provided. Replace the other plastic cap on the outer conductor and store the center and outer conductors
in the foam lined storage case.
85051B
Use, Maintenance, and Care of the Devices
Gaging Connectors