AC Power Supply, 12VDC Cable, Power Cord and Outer Case:
Turn OFF the eQuinox and disconnect from AC or DC power
before any cleaning or disinfection activity. DO NOT spray the
outer case directly. Use a damp (not wet) cloth or sponge. Spray
the cloth or sponge with a mild detergent solution to clean the
cabinet and power supplies. To disinfect the eQuinox, use Lysol®
Brand II disinfectant (or equivalent). Proceed as directed by the
cleaner manufacturer.
WARNING: ELECTRICAL SHOCK HAZARD. TURN OFF THE
UNIT AND DISCONNECT THE POWER CORD FROM THE
ELECTRICAL OUTLET BEFORE YOU CLEAN THE UNIT TO
PREVENT ACCIDENTAL ELECTRICAL SHOCK AND BURN HAZARD. ONLY
YOUR EQUIPMENT PROVIDER OR A QUALIFIED SERVICE TECHNICIAN
SHOULD REMOVE THE COVERS OR SERVICE THE UNIT. DO NOT USE
LIQUID DIRECTLY ON THE UNIT. A LIST OF UNDESIRABLE CHEMICAL
AGENTS INCLUDES BUT IS NOT LIMITED TO THE FOLLOWING: AL-
COHOL AND ALCOHOL-BASED PRODUCTS, CONCENTRATED CHLO-
RINE-BASED PRODUCTS (ETHYLENE CHLORIDE), AND OIL- BASED
PRODUCTS (PINE-SOL®, LESTOIL®). THESE ARE NOT TO BE USED TO
CLEAN THE PLASTIC HOUSING ON OXYGEN CONCENTRATOR, AS THEY
CAN DAMAGE THE UNIT'S PLASTIC.
WARNING: ELECTRICAL SHOCK HAZARD. DISCONNECT THE POWER
CORD FROM THE ELECTRICAL OUTLET BEFORE YOU CLEAN THE UNIT
TO PREVENT ACCIDENTAL ELECTRICAL SHOCK AND BURN HAZARD.
ONLY YOUR EQUIPMENT PROVIDER OR A QUALIFIED SERVICE TECHNI-
CIAN SHOULD REMOVE THE COVERS OR SERVICE THE UNIT.
WARNING: CARE SHOULD BE TAKEN TO PREVENT THE OXYGEN CON-
CENTRATOR AND THE BATTERY FROM GETTING WET OR ALLOWING
FLUIDS TO ENTER THE UNIT. THIS CAN CAUSE A MALFUNCTION OR
SHUT DOWN, AND CAUSE AN INCREASED RISK FOR ELECTRICAL
SHOCK OR BURNS.
WARNING: CLEAN THE CABINET, CONTROL PANEL, AND POWER
CORD ONLY WITH A MILD HOUSEHOLD CLEANER APPLIED WITH A
DAMP (NOT WET) CLOTH OR SPONGE, AND THEN WIPE ALL SURFACES
DRY. DO NOT ALLOW ANY LIQUID TO GET INSIDE THE DEVICE. PAY
SPECIAL ATTENTION TO THE OXYGEN OUTLET FOR THE CANNULA
CONNECTION TO MAKE SURE IT REMAINS FREE OF DUST, WATER,
AND PARTICLES.
18 — 4805-C3_01 A
WARNING: DO NOT USE LIQUID DIRECTLY ON THE UNIT.
A LIST OF UNDESIRABLE CHEMICAL AGENTS INCLUDES
BUT IS NOT LIMITED TO THE FOLLOWING: ALCOHOL AND
ALCOHOL-BASED PRODUCTS, CONCENTRATED CHLORINE-BASED
PRODUCTS (ETHYLENE CHLORIDE), AND OIL- BASED PRODUCTS
(PINE-SOL®, LESTOIL®). THESE ARE NOT TO BE USED TO CLEAN THE
PLASTIC HOUSING ON OXYGEN CONCENTRATOR, AS THEY CAN DAM-
AGE THE UNIT'S PLASTIC.
Optional Humidifier
contamination. Follow the cleaning recommendations provided
CAUTION: DO NOT operate the eQuinox and ambulate
while a humidifier is attached. Remove the humidifier
bottle before walking. DO NOT lay the eQuinox down while
attached to a humidifier bottle.
Universal Cart
Before cleaning the Universal Cart, remove it from the eQuinox.
Use a damp (not wet) cloth or sponge and mild detergent solution
to clean the Universal Cart.
Power Cartridge (Battery)
The Battery in the eQuinox requires special care to assure a
longer life and the highest level of performance. The SeQual
Battery is the only approved Battery recommended for use with
the eQuinox.
Use a damp (not wet) cloth or sponge to clean the Battery. First
spray the cloth or sponge with a mild detergent then clean the
Battery case and the latch.
SeQual Recommended Monthly Battery Service: Run the
eQuinox on Battery completely down to power loss condition
(red light and alarm). Plug eQuinox into AC power and recharge
until full on indicator. The device does not have to be running to
charge (which allows for shorter charge time).
SeQual recommends replacing the power cartridge after 2 years.
Battery Disposal
Your Battery is rechargeable and can be recycled. Always return
it to your home care provider for proper disposal. You can also
disposal of the Battery.