CAIRE eQuinox User Manual page 13

Table of Contents

Advertisement

Typical New Power Cartridge Operating Times
Operating Type
Continous Mode
1 LPM
2 LPM
3 LPM
Pulse Mode (12 BPM)
Bolus 16 mL
Bolus 48 mL
Bolus 96 mL
Note: Operating the Oxygen Concentrator outside of its normal operating
temperature range can affect performance and decrease battery run time
and/or increase battery charge time. (Refer to the Specifications section in
this manual.) You will achieve longer operating time on the Battery if you
operate your eQuinox in the Pulse Mode (Physician approved mode).
Battery Type
12-cell
24-cell
1.96 hrs
4.4 hrs
1.1 hrs
2.5 hrs
0.73 (44 min)
1.67 hrs
2.38 hrs
5.4 hrs
2.16 hrs
4.9 hrs
1.54 hrs
3.5 hrs
Typical Battery Recharge Time
The typical time to recharge your Battery to achieve 80%
capacity from a fully discharged Battery is between 2 and 5
is fully charged, the battery gauge will cease cascading. The
battery can be charged in the unit when the eQuinox is turned off
and plugged into the portable oxygen concentrator.
If the Battery becomes too warm during discharging, recharging
Battery and allowing it to cool may expedite this cooling process.
The Battery will not recharge when operating from the 12VDC
Cable.
The Battery charges when in the device is operating on AC power
only. The Battery will also charge from the optional Desktop
Charger.
eQuinox
eQuinox User Manual — 13

Advertisement

Table of Contents
loading

Table of Contents