12VDC Cable
12VDC Cable
A 12VDC Cable allows the system to operate from DC outlets,
such as those found in motor vehicles.
1. Start your vehicle.
2. Insert the 12 Volt DC cord into the recessed power
receptacle on the side of the device.
3. Insert the 12VDC Cable plug into the DC power outlet in
your motor vehicle.
When the device is properly connected and receiving power from
the DC power source, a green indicator light on the Power Supply
and the External Power Indicator
illuminate.
SeQual eQuinox DC Supply Connection
When the eQuinox is connected to the 12VDC Cable in a vehicle,
the following applies:
• The eQuinox will run at 3.0LPM on continuous flow while
being operated with the DC Supply in a car.
• The eQuinox will charge the battery while on 12VDC Cable in
a car while running on continuos flow at 2LPM or less.
• If the eQuinox battery is being charged, the battery symbol will
waterfall. If the eQuinox battery is not charging, the battery
symbol will be displayed solid.
WARNING: DO NOT ALLOW EITHER THE AIR INTAKE OR THE
AIR OUTLET VENTS TO BECOME BLOCKED. THIS CAN CAUSE
THE OXYGEN CONCENTRATOR TO OVERHEAT AND IMPAIR
PERFORMANCE.
WARNING: IF THE VEHICLE'S DC POWER SOURCE DROPS BELOW 11.5
VOLTS, THE EQUINOX WILL REVERT TO THE BATTERY OPERATION (IF
THE BATTERY IS PRESENT AND CHARGED). (THE EXTERNAL POWER
INDICATOR MAY REMAIN ON.)
on the control panel will
Note: While operating the device from a 12VDC Cable, the Battery may
not charge.
Note: When operating the eQuinox from a 12VDC Cable, it is recommend-
ed that the Pulse Dose Mode be used if prescribed to use a pulse dose
device.
Note: To ensure that the device is utilizing the DC power, check that the
External Power Indicator is illuminated. If a beep is heard, this could indi-
cate a loss of external power. Check if the green Battery Power Verification
LED is on or off. If the battery is removed on DC power, it will ensure no
loss of charge while the unit is running. If the indicator is not illuminated,
the battery is powering the device.
Settings Available When Operating on DC for eQuinox
Flow Setting
Continuous
0.5
Yes
1.0
Yes
1.5
Yes
2.0
Yes
2.5
Yes
3.0
Yes
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
Note: Actual performance of the eQuinox may vary depending upon usage
conditions of the motor vehicle.
eQuinox
Bolus Size (mL)
Pulse
-
-
16
Yes
24
Yes
32
Yes
40
Yes
48
Yes
56
Yes
64
Yes
72
Yes
80
Yes
88
Yes
96
Yes
128
Yes
160
Yes
192
Yes
eQuinox User Manual — 11