Electrical Specifications (ambient temperature 23°C ± 5°C)
Light meter
Measurement
LUX
(auto range)
Foot-candles
(auto range)
Range
2,000 LUX
20,000 LUX
100,000 LUX
200 Ft-cd
2,000 Ft-cd
10,000 Ft-cd
Note: Accuracy tested using a standard tungsten lamp (2856 K degree temperature)
Temperature meter
Sensor Type
Type K
Type J
All rights reserved including the right of reproduction in whole or in part in any form
www.ponpe.com
Range
2,000LUX
20,000 LUX
100,000 LUX
200Ft-cd
2,000 Ft-cd
10,000 Ft-cd
Resolution
1 LUX
10 LUX
100 LUX
0.1 Ft-cd
1 Ft-cd
10 Ft-cd
Resolution
0.1 °C
0.1 °F
0.1 °C
0.1 °F
Copyright © 2014 FLIR Systems, Inc.
www.extech.com
12
Actual Display for each range
0 to 1,999 LUX
1,800 to 19,990 LUX
18,000 to 99,900 LUX
0 to 186.0 Ft-cd
167.0 to 1,860.0 Ft-cd
1,670.0 to 9,290.7 Ft-cd
Accuracy
± (4%rdg + 2 dgt)
± (4%rdg + 2 Ft-cd)
± (4%rdg + 20 Ft-cd)
Range
-50.0 to 1300.0°C
-50.1 to -100.0°C
-58.0 to 2372.0°F
-58.1 to -148.0°F
-50.0 to 1200.0°C
-50.1 to -100.0°C
-58.0 to 2192.0°F
-58.1 to -148.0°F
Accuracy
± (0.4%rdg + 0.5°C)
± (0.4%rdg + 1.0°C)
± (0.4%rdg + 1.0°F)
± (0.4%rdg + 1.8°F)
± (0.4%rdg + 0.5°C)
± (0.4%rdg + 1.0°C)
± (0.4rdg% + 1.0°F)
± (0.4rdg% + 1.8°F)
SDL400-en-GB_V1.6 12/14
Contact
sale@ponpe.com
Need help?
Do you have a question about the SDL400 and is the answer not in the manual?
Questions and answers