10. FUNCTION SETTING OF THE BALANCE
10-1 Display Lock Function
To use this function, set "Display lock function ( ploc )" to "ON ( 1 )".
Why use this function
This function is used to compensate time-elapsed errors caused by evaporation.
Even when the evaporation trap is used, an evaporation of 0.05 mg per minute (0.05 L per minute)
will occur. This evaporation influences the measurement, particularly when the volume of the test
liquid is small. For such a situation, the display lock function is useful.
How this function works
After the test liquid is dispensed from the pipette, measurement starts. When the mass value is
stable, data is averaged. After a certain period of time, the averaged value is locked in the display.
Processing indicator
Blinking
Averaging process
NOTE
The RE-ZERO operation unlocks the display.
10-2 Auto Print Mode
In the balance function setting of "Data output ( dout )", when "Data output mode ( prt )" is set to
"Auto print mode A ( 1 )", the mass value will be transmitted to WinCT-Pipette automatically once the
mass value of the dispensed test liquid becomes stable.
Processing indicator
Stabilization indicator
Illuminated
Locked display
20