9-1 Measuring Procedure
Using the predetermined pipette volume, perform the measurement the specified times.
ISO 8655 recommends ten measurements for each for the volumes described below.
At the nominal volume of a fixed volume pipette
At the following three points of an adjustable volume pipette
1. 10% of the maximum volume or the lower limit of the adjustable volume range, whichever is greater.
2. 50% of the maximum volume
3. 100% of the maximum volume
NOTE
Pipette volumes and number of measurements can be changed.
9-2 Judging the Results
WinCT-Pipette converts the mass values into volumetric values, compares these with the
predetermined judgment standard and performs judgment whether the pipette will pass or fail.
Judgment standard
The averaged value (exceeds / does not exceed) the specified maximum permissible error.
Repeatability (exceeds / does not exceed) the specified maximum permissible error.
9-3 Example Causes of Measurement Errors
Evaporation of the test liquid
Evaporation of the distilled water dispensed from the pipette into the sample cup can be a cause of
measurement error.
To reduce errors caused by evaporation
Be sure to use the evaporation trap. Using the evaporation trap, evaporation of the test liquid can
be reduced to 0.05 mg per minute (0.05 L per minute).
Use the display lock function of the BM series. For details, refer to "10-1 Display Lock Function".
Water temperature and barometric pressure
Correctness of the Z factor, that is used to convert a mass value to a volumetric value, depends on
the density by the water temperature.
The barometric pressure, even if changed in the range between 950 hPa and 1050 hPa, influences
measurements only by 0.01%.
Water temperature, if changed by 5°C, influences measurements by 0.11%.
9. MEASUREMENT
18