and opening the cylinder again. If this does not help, shut the system down again and contact a
service representative.
5.
Adjust the flow rate setting by turning the knob clockwise until the desired setting shows through the
window.
WARNING: NEVER try to remove the Gas Regulator unless the cylinder is off and pressure has been
relieved.
WARNING: If at any time the Gas Regulator, cylinder or other apparatus does not operate in its usual or
normal manner, stop using the apparatus and notify your dealer immediately.
WARNING: There is no flow between increments.
ATTENTION: Flow rates are not accurate when a back-pressure of more than 2 PSIG exists downstream of
the Gas Regulator.

Closing The Regulator

1.
When you have finished using oxygen, close the system by turning the cylinder valve clockwise to the
"OFF" position.
CAUTION:
Oxygen will continue to flow from the outlet orifice assembly until the cylinder valve is
closed (Reference CGA Pamphlet P-14, "Accident Prevention in Oxygen Rich and Oxygen
Deficient Atmospheres"). It is very important to always close the cylinder valve when not in
use.

Removing The Regulator From A Cylinder

1.
Close the cylinder valve.
2.
Wait for the oxygen to flow out of the Gas Regulator, then turn the knob to the "OFF" position
3.
Remove the Gas Regulator from the cylinder by loosening the T-handle or removing the inlet nut.
4.
When not attached to a cylinder, store the Gas Regulator in its original container in a clean protected
area free from grease, oil and other contamination.
5.
Amvex recommends replacing the sealing washer on Yoke style inlet connection when the Gas
Regulator is mounted to a different cylinder.

CLEANING INSTRUCTIONS

Clean the external surfaces of the Gas Regulator with a damp cloth. If more cleaning is required, return the
Gas Regulator to your dealer.
CAUTION:
DO NOT submerse the Gas Regulator in any form of liquid. This will cause damage and
void any warranty on the product.