CONSTANT SUCTION UNIT OPERATOR/MAINTENANCE Ohio Medical Corporation 1111 Lakeside Drive Gurnee, Illinois 60031-4099 1-800-662-5822/847-855-0800 Fax: 847/855-6218 www.ohiomedical.com...
Page 2
If, after reading this manual you require additional information, please contact Ohio Medicals’ Customer Service Department at 800-662-5822 or your local Ohio Medical distributor. Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218 www.ohiomedical.com —– 550107 (Rev.5) 08/2005...
9. Place a collection canister bracket (ring) in the slide mounting bracket of the backpole assembly or pedestal stand and insert a collection canister. Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218 www.ohiomedical.com —– 550107 (Rev.5) 08/2005...
Page 4
Place the Chest Drainage Unit in the “optional” Ohio Medical Chest Drainage Unit Bracket. Ensure that tubing is connected from the “VACUUM PORT ASSEMBLY” of the Moblvac III-cs to the Ohio Medical Needle Valve Assembly which is mounted in the collection canister bracket of the backpole.
Page 5
Chest Drainage Units With Built In Flow Adjustment Valves. Place the Chest Drainage Unit in the “optional” Ohio Medical Chest Drainage Unit Bracket. Ensure that tubing is connected from the “VACUUM PORT ASSEMBLY” of the Moblvac III-cs to the vac- ®...
Tighten the set screw until regulator knob is secure to shaft and cannot be pulled off by hand. Reconnect all tubing. VACUUM REGULATOR CLEANING Remove the vacuum regulator assembly as described in the VACUUM REGULATOR RE- PLACEMENT section above. Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218 www.ohiomedical.com —– 550107 (Rev.5) 08/2005...
However, if performance has been affected by the pump becoming contaminated with aspirant, or if an overflow problem has occurred, the following Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218 www.ohiomedical.com —– 550107 (Rev.5) 08/2005...
Page 9
Remove the pump assembly from the Moblvac III-cs as described above in the PUMP REPLACEMENT section. .Remove the three (3) pump head screws located on the pump head. Remove the cover plate, shim, wear plate and the four (4) vanes.
3. Loose fittings on the exhaust side of 3. Check all fittings on the exhaust side the pump. of the pump for any loose connection. Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218 www.ohiomedical.com —– 550107 (Rev.5) 08/2005...
Page 11
FM 33489 MEDICAL EQUIPMENT LISTED 544U TO ELECTRIC SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 2601-1, AND C22.2 NO. 601.1 #9N56 Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218 www.ohiomedical.com —– 550107 (Rev.5) 08/2005...