INSTRUCTIONS FOR USE GAS REGULATORS – CLICK STYLE AND MINI CLICK STYLE Rx Only Basic matrix shown. Consult the Ohio Medical website for full matrix or contact your Ohio Medical representative. The product package label contains your product identifier and description.
Page 2
IMPORTANT: SAFETY INSTRUCTIONS This manual provides you with important information about the Gas Regulator and should be read carefully to ensure the safe and proper use of this product. Read and understand all the safety and operating instructions contained in this booklet before using this product.
The Gas Regulators are available for Nut and Nipple, Yoke style CGA and Bullnose fittings. The units are machined brass finished with an anodized aluminum body. The Amvex Click Style Gas Regulators are available in a variety of flow rates from 0-4, 0-8, 0-15, 0-25 L/min. SPECIFICATIONS Gas Regulator Description 1.
OPERATING INSTRUCTIONS Inspect the gas cylinder valve for leaks before putting it into service. If a leak is found around the cylinder valve stem, close the valve, place the cylinder outdoors in a safe area and inform your gas supplier immediately.
and opening the cylinder again. If this does not help, shut the system down again and contact a service representative. Adjust the flow rate setting by turning the knob clockwise until the desired setting shows through the window. WARNING: NEVER try to remove the Gas Regulator unless the cylinder is off and pressure has been relieved.